Controlled clinical comparison of bioMérieux VITAL and BACTEC NR-660 blood culture systems for detection of bacteremia and fungemia in adults.

M L Wilson1, S Mirrett, L C McDonald, M P Weinstein, J Fune, L B Reller.   

Abstract

A total of 9,446 blood cultures were collected from adult patients at three university-affiliated hospitals. Of these, 8,943 cultures were received with both aerobic bottles filled adequately; 885 yielded 1,016 microorganisms, including 622 isolates (61%) that were the cause of sepsis, 337 isolates (33%) that were contaminants, and 57 isolates (6%) that were indeterminate as the cause of sepsis. With the exception of Staphylococcus aureus, which was recovered more often from VITAL aerobic bottles, more pathogenic microorganisms were recovered from BACTEC NR6 (aerobic) bottles than from VITAL aerobic bottles. Growth of pathogenic microorganisms was detected earlier in VITAL aerobic bottles. A total of 8,647 blood cultures were received with both anaerobic bottles filled adequately; 655 yielded 740 microorganisms, including 486 isolates (66%) that were the cause of sepsis, 215 isolates (29%) that were contaminants, and 39 isolates (6%) that were indeterminate as the cause of sepsis. More pathogenic microorganisms were recovered from VITAL anaerobic bottles than from BACTEC NR7 (anaerobic) bottles. Growth of pathogenic microorganisms was detected earlier in VITAL anaerobic bottles. In 8,500 sets all four bottles were received adequately filled. When paired aerobic and anaerobic bottle sets (systems) were compared, more pathogenic microorganisms (again with the exception of S. aureus) were recovered from the BACTEC system. For the 304 septic episodes (253 unimicrobial and 51 polymicrobial), significantly more were detected by the BACTEC system. We conclude that VITAL requires modification to improve recovery of pathogenic microorganisms to make it competitive with other commercially available blood culture systems.

Entities:  

Mesh:

Year:  1999        PMID: 10325312      PMCID: PMC84930     

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  15 in total

1.  Effects of atmosphere of incubation and of routine subcultures on detection of bacteremia in vacuum blood culture bottles.

Authors:  J L Harkness; M Hall; D M Ilstrup; J A Washington
Journal:  J Clin Microbiol       Date:  1975-10       Impact factor: 5.948

2.  Clinical comparison of a new automated infrared blood culture system with the BACTEC 460 system.

Authors:  D Jungkind; J Millan; S Allen; J Dyke; E Hill
Journal:  J Clin Microbiol       Date:  1986-02       Impact factor: 5.948

3.  Diminished growth of Pseudomonas aeruginosa in unvented blood-culture bottles.

Authors:  J G Knepper; B F Anthony
Journal:  Lancet       Date:  1973-08-11       Impact factor: 79.321

4.  Detection kinetics for positive blood culture bottles by using the VITAL automated system.

Authors:  H Marchandin; B Compan; M Simeon De Buochberg; E Despaux; C Perez
Journal:  J Clin Microbiol       Date:  1995-08       Impact factor: 5.948

5.  Detection of yeasts and filamentous fungi in blood cultures during a 10-year period (1972 to 1981).

Authors:  J Bille; L Stockman; G D Roberts
Journal:  J Clin Microbiol       Date:  1982-11       Impact factor: 5.948

6.  The clinical significance of positive blood cultures: a comprehensive analysis of 500 episodes of bacteremia and fungemia in adults. I. Laboratory and epidemiologic observations.

Authors:  M P Weinstein; L B Reller; J R Murphy; K A Lichtenstein
Journal:  Rev Infect Dis       Date:  1983 Jan-Feb

7.  Controlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems.

Authors:  M L Wilson; M P Weinstein; L G Reimer; S Mirrett; L B Reller
Journal:  J Clin Microbiol       Date:  1992-02       Impact factor: 5.948

8.  Multicenter comparison of the high volume (10 ml) NR BACTEC PLUS system and the standard (5 ml) NR BACTEC system.

Authors:  F P Koontz; K K Flint; J K Reynolds; S D Allen
Journal:  Diagn Microbiol Infect Dis       Date:  1991 Mar-Apr       Impact factor: 2.803

9.  Controlled evaluation of BACTEC PLUS 27 and Roche Septi-Chek anaerobic blood culture bottles.

Authors:  M L Wilson; L J Harrell; S Mirrett; M P Weinstein; C W Stratton; L B Reller
Journal:  J Clin Microbiol       Date:  1992-01       Impact factor: 5.948

10.  Clinical comparison of an agar slide blood culture bottle with tryptic soy broth and a conventional blood culture bottle with supplemented peptone broth.

Authors:  M P Weinstein; L B Reller; S Mirrett; W L Wang; D V Alcid
Journal:  J Clin Microbiol       Date:  1985-05       Impact factor: 5.948

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.