Literature DB >> 10204885

Analytic approach to the design of transverse gradient coils with co-axial return paths.

R Bowtell1, A Peters.   

Abstract

Transverse gradient coils with co-axial return paths offer reduced acoustic noise compared with standard cylindrical gradient coils, due to local force balancing, and can also easily be made to have a length to diameter ratio that is less than one. Analytic expressions for the magnetic field and vector potential generated by this type of coil are described here, along with a formula for calculating the coil inductance. It is shown that these expressions allow the implementation of powerful analytic methods of coil design, as well as the incorporation of active magnetic screening. It is also demonstrated how the mathematics specifies the best parameters to use when designing coils with small numbers of elements. A head gradient coil for use at 3.0 T has been designed using the analytic approach described here. The process of coil design and construction is outlined and the performance of the coil in comparison with a similar standard cylindrical coil is described.

Mesh:

Year:  1999        PMID: 10204885     DOI: 10.1002/(sici)1522-2594(199903)41:3<600::aid-mrm24>3.0.co;2-e

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  6 in total

1.  Acoustic noise during functional magnetic resonance imaging.

Authors:  M E Ravicz; J R Melcher; N Y Kiang
Journal:  J Acoust Soc Am       Date:  2000-10       Impact factor: 1.840

Review 2.  Acoustic noise concerns in functional magnetic resonance imaging.

Authors:  Adriaan Moelker; Peter M T Pattynama
Journal:  Hum Brain Mapp       Date:  2003-11       Impact factor: 5.038

3.  New head gradient coil design and construction techniques.

Authors:  William B Handler; Chad T Harris; Timothy J Scholl; Dennis L Parker; K Craig Goodrich; Brian Dalrymple; Frank Van Sass; Blaine A Chronik
Journal:  J Magn Reson Imaging       Date:  2013-10-09       Impact factor: 4.813

4.  High slew-rate head-only gradient for improving distortion in echo planar imaging: Preliminary experience.

Authors:  Ek T Tan; Seung-Kyun Lee; Paul T Weavers; Dominic Graziani; Joseph E Piel; Yunhong Shu; John Huston; Matt A Bernstein; Thomas K F Foo
Journal:  J Magn Reson Imaging       Date:  2016-02-26       Impact factor: 4.813

5.  Lightweight, compact, and high-performance 3T MR system for imaging the brain and extremities.

Authors:  Thomas K F Foo; Evangelos Laskaris; Mark Vermilyea; Minfeng Xu; Paul Thompson; Gene Conte; Christopher Van Epps; Christopher Immer; Seung-Kyun Lee; Ek T Tan; Dominic Graziani; Jean-Baptise Mathieu; Christopher J Hardy; John F Schenck; Eric Fiveland; Wolfgang Stautner; Justin Ricci; Joseph Piel; Keith Park; Yihe Hua; Ye Bai; Alex Kagan; David Stanley; Paul T Weavers; Erin Gray; Yunhong Shu; Matthew A Frick; Norbert G Campeau; Joshua Trzasko; John Huston; Matt A Bernstein
Journal:  Magn Reson Med       Date:  2018-03-13       Impact factor: 4.668

6.  Peripheral nerve stimulation characteristics of an asymmetric head-only gradient coil compatible with a high-channel-count receiver array.

Authors:  Seung-Kyun Lee; Jean-Baptiste Mathieu; Dominic Graziani; Joseph Piel; Eric Budesheim; Eric Fiveland; Christopher J Hardy; Ek Tsoon Tan; Bruce Amm; Thomas K-F Foo; Matt A Bernstein; John Huston; Yunhong Shu; John F Schenck
Journal:  Magn Reson Med       Date:  2015-12-02       Impact factor: 4.668

  6 in total