Assessment of image quality in dental radiography, part 1: phantom validity.

K Yoshiura1, T Kawazu, T Chikui, M Tatsumi, K Tokumori, T Tanaka, S Kanda.   

Abstract

OBJECTIVE: The purpose of this study was to describe and validate an image-quality phantom to be used in dental radiography for comparison of film and digitally acquired images. STUDY
DESIGN: An aluminum block of 12 steps, with 7 holes in each step, was covered by acrylic blocks. This phantom was radiographed with Kodak Ultra-speed and Ektaspeed Plus films at 70, 65, and 60 kVp with the whole exposure range available. All together, 50 dental films were randomly sequenced and presented to 7 observers. The average number of perceptible holes from all steps was plotted against exposure for each tube voltage and film type, generating a modified perceptibility curve. The tentative optimum exposure level was determined from perceptibility curves in each experimental condition and compared with that determined by means of the standard aluminum stepwedge and the preset time of the x-ray machine. The density range of this phantom at the optimum exposure was compared with that of clinical dental radiographs. Validity of the phantom was evaluated according to the optimum exposure level from the modified perceptibility curves and the overall density range. Finally, the average maximum numbers of perceptible holes at the tentative optimum exposure level were compared for each tube voltage and film type. The statistical test used was a 2-way factorial analysis of variance.
RESULTS: The exposure at the perceptibility curve peak approximated that obtained by means of the standard aluminum step-wedge and the time preset by the manufacturer. The overall density range at the perceptibility curve peak covered the clinical density range for each tube voltage and film type. There were no statistically significant differences between film types or among tube voltages.
CONCLUSIONS: The x-ray attenuation range for this phantom seemed to approximate clinical conditions. In addition, differences in image quality could be quantitatively evaluated by means of the number of the holes seen in the phantom.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 9927090     DOI: 10.1016/s1079-2104(99)70304-5

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  8 in total

1.  Comparison of psychophysical properties of two intraoral digital sensors on low-contrast perceptibility.

Authors:  X-Q Shi; D Benchimol; K Näsström
Journal:  Dentomaxillofac Radiol       Date:  2013-10-29       Impact factor: 2.419

2.  A new method to evaluate image quality of CBCT images quantitatively without observers.

Authors:  Yohei Takeshita; Mayumi Shimizu; Kazutoshi Okamura; Shoko Yoshida; Warangkana Weerawanich; Kenji Tokumori; Gainer R Jasa; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

3.  A mathematical model relating changes of grey values to changes of thicknesses of a stepwedge.

Authors:  H Geha; B Bechara; T Faddoul; M Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2012-07-27       Impact factor: 2.419

4.  Prediction of detectability of the mandibular canal by quantitative image quality evaluation using cone beam CT.

Authors:  Yohei Takeshita; Mayumi Shimizu; Gainer R Jasa; Warangkana Weerawanich; Kazutoshi Okamura; Shoko Yoshida; Kenji Tokumori; Junichi Asaumi; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2018-02-13       Impact factor: 2.419

Review 5.  The missing link in image quality assessment in digital dental radiography.

Authors:  Kazutoshi Okamura; Kazunori Yoshiura
Journal:  Oral Radiol       Date:  2019-07-13       Impact factor: 1.852

6.  Image quality evaluation of eight complementary metal-oxide semiconductor intraoral digital X-ray sensors.

Authors:  Sorin Teich; Wisam Al-Rawi; Masahiro Heima; Fady F Faddoul; Gil Goldzweig; Zvi Gutmacher; Dror Aizenbud
Journal:  Int Dent J       Date:  2016-04-22       Impact factor: 2.607

7.  A Novel Phantom and a Dedicated Developed Software for Image Quality Controls in X-Ray Intraoral Devices.

Authors:  Giuseppe Acri; Sergio Gurgone; Claudio Iovane; Marco B Romeo; Daniele Borzelli; Barbara Testagrossa
Journal:  J Biomed Phys Eng       Date:  2021-04-01

8.  Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation.

Authors:  Taku Kuramoto; Shinya Takarabe; Kazutoshi Okamura; Kenshi Shiotsuki; Yusuke Shibayama; Hiroki Tsuru; Hiroshi Akamine; Masato Tatsumi; Toyoyuki Kato; Junji Morishita; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2020-05-07       Impact factor: 2.419

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.