A cone-beam megavoltage CT scanner for treatment verification in conformal radiotherapy.

M A Mosleh-Shirazi1, P M Evans, W Swindell, S Webb, M Partridge.   

Abstract

PURPOSE: A prototype scanner for large-volume megavoltage computed tomography (MVCT) in a clinical set-up is described. The ultimate aim is to improve treatment accuracy in conformal radiotherapy through patient set-up error reduction and transit dosimetry.
MATERIALS AND METHODS: The scanner consists of a custom-built 2D CsI(Tl) crystal array viewed by a lens and a CCD camera. Image acquisition is synchronized with radiation pulses. The 2D projections resulting from a single continuous 360 degrees gantry rotation are reconstructed using a cone-beam tomography algorithm. Prior to reconstruction, the raw projections are calibrated and corrected for centre of rotation movement and accelerator output fluctuation. The performance of the system has been evaluated by reconstructing projections of open fields, test objects and a humanoid phantom.
RESULTS: Hundreds of 2D projections can be acquired with a clinically-acceptable data collection time (about 2 min) and dose (approximately 40 cGy, with a possible four-fold reduction). A maximum density resolution of about 2% is achieved offering some soft tissue discrimination without using image enhancement tools. A spatial resolution of 2.5 mm is obtained. The reconstructed image intensity is linear with electron density over the range of interest. Coronal or sagittal slices through the 3D reconstruction of the humanoid phantom show a better delineation of structures than the corresponding portal images taken at the same orientation.
CONCLUSIONS: A similar image quality to our current single-slice MVCT scanner is achieved with the advantage of providing tens of tomographic slices for a single gantry rotation. This work demonstrates the feasibility of clinical cone-beam MVCT and indicates how this prototype can be improved.

Entities:  

Mesh:

Year:  1998        PMID: 9925252     DOI: 10.1016/s0167-8140(98)00042-5

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  15 in total

1.  Accuracy estimation for projection-to-volume targeting during rotational therapy: a feasibility study.

Authors:  Yong Long; Jeffrey A Fessler; James M Balter
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

3.  High-DQE EPIDs based on thick, segmented BGO and CsI:Tl scintillators: performance evaluation at extremely low dose.

Authors:  Yi Wang; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Louis Perna
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

4.  Accuracy in automatic image registration between MV cone beam computed tomography and planning kV computed tomography in image guided radiotherapy.

Authors:  Nithya Kanakavelu; E James Jebaseelan Samuel
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-22

5.  A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance.

Authors:  Chiaho Hua; Weiguang Yao; Takao Kidani; Kazuo Tomida; Saori Ozawa; Takenori Nishimura; Tatsuya Fujisawa; Ryousuke Shinagawa; Thomas E Merchant
Journal:  Br J Radiol       Date:  2017-10-09       Impact factor: 3.039

6.  A survey of image-guided radiation therapy use in the United States.

Authors:  Daniel R Simpson; Joshua D Lawson; Sameer K Nath; Brent S Rose; Arno J Mundt; Loren K Mell
Journal:  Cancer       Date:  2010-08-15       Impact factor: 6.860

7.  Characterization of linear accelerator X-ray source size using a laminated beam-spot camera.

Authors:  Collins Yeboah
Journal:  J Appl Clin Med Phys       Date:  2011-05-10       Impact factor: 2.102

8.  Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate.

Authors:  Douglas J Moseley; Elizabeth A White; Kirsty L Wiltshire; Tara Rosewall; Michael B Sharpe; Jeffrey H Siewerdsen; Jean-Pierre Bissonnette; Mary Gospodarowicz; Padraig Warde; Charles N Catton; David A Jaffray
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-03-01       Impact factor: 7.038

9.  Current concepts on imaging in radiotherapy.

Authors:  Michela Lecchi; Piero Fossati; Federica Elisei; Roberto Orecchia; Giovanni Lucignani
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-10-31       Impact factor: 9.236

10.  Modelling the transport of optical photons in scintillation detectors for diagnostic and radiotherapy imaging.

Authors:  Emilie Roncali; Mohammad Amin Mosleh-Shirazi; Aldo Badano
Journal:  Phys Med Biol       Date:  2017-10-04       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.