Magnetic resonance imaging with implanted neurostimulators: an in vitro and in vivo study.

V M Tronnier1, A Staubert, S Hähnel, A Sarem-Aslani.   

Abstract

OBJECTIVE: The goal was to assess the safety of magnetic resonance imaging (MRI) with implanted neurostimulators, in an in vitro and in vivo study.
METHODS: Two different implantable pulse generators (IPGs) (ITREL II and 3; Medtronic, Minneapolis, MN) and different leads (separately and connected to an IPG) were tested in three different MRI scanners (0.2, 0.25, and 1.5 T). Measurements of the induced voltages (using an external oscilloscope) and the induced heat (using an infrared camera) were performed in an in vitro study. Finally, 38 patients with implanted neurostimulator systems (leads and IPGs) underwent MRI in 50 examinations, with continuous monitoring by a physician with uninterrupted visual and vocal contact with the patient. Twenty-five patients were studied prospectively, with documented printouts of the parameter settings before and after MRI.
RESULTS: An induced voltage of 2.4 to 5.5 V was measured in the experimental configuration with a lead connected to an IPG. The voltage was higher with the leads alone, compared with the leads connected to the IPG, and was dependent on the MRI scanner, the sequences, and the type of lead. No heat induction was observed in any part of the hardware. No change of pulse shape or change of IPG parameters was observed during MRI. No adverse effects occurred in patients with chronically implanted deep brain leads connected to an IPG.
CONCLUSION: MRI can be safely performed in patients with implanted neurostimulation systems with the tested deep brain leads connected to an IPG (ITREL II and 3), with running parameters. No heat induction was detected, and the experimentally measured induced voltage did not seem to harm the patients. Only the reed switch of the IPGs was activated; the other parameters remained unchanged. Further investigations must be performed to study the local electrical effects in larger plate electrodes; these effects might cause slight discomfort. There is no danger with any type of electrode during MRI examinations if the electrodes lie outside the region of interest. These observations are restricted to the tested devices. A conscientious estimation of the risks and benefits of MRI for patients with implanted devices is recommended. If the type of device is not known to the examiner, MRI should still be considered to be contraindicated.

Entities:  

Mesh:

Year:  1999        PMID: 9894972     DOI: 10.1097/00006123-199901000-00073

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  13 in total

1.  [Recurrent disc herniation during chronic pain therapy].

Authors:  M Lorenz
Journal:  Schmerz       Date:  2006-08       Impact factor: 1.107

2.  Targeting the subthalamic nucleus for deep brain stimulation: technical approach and fusion of pre- and postoperative MR images to define accuracy of lead placement.

Authors:  N A Hamid; R D Mitchell; P Mocroft; G W M Westby; J Milner; H Pall
Journal:  J Neurol Neurosurg Psychiatry       Date:  2005-03       Impact factor: 10.154

Review 3.  MRI-related heating near deep brain stimulation electrodes: more data are needed.

Authors:  Akshay A Gupte; Devashish Shrivastava; Maggie A Spaniol; Aviva Abosch
Journal:  Stereotact Funct Neurosurg       Date:  2011-04-13       Impact factor: 1.875

4.  Cortical magnetoencephalography of deep brain stimulation for the treatment of postural tremor.

Authors:  Allison T Connolly; Jawad A Bajwa; Matthew D Johnson
Journal:  Brain Stimul       Date:  2012-02-22       Impact factor: 8.955

Review 5.  Improving Safety of MRI in Patients with Deep Brain Stimulation Devices.

Authors:  Alexandre Boutet; Clement T Chow; Keshav Narang; Gavin J B Elias; Clemens Neudorfer; Jürgen Germann; Manish Ranjan; Aaron Loh; Alastair J Martin; Walter Kucharczyk; Christopher J Steele; Ileana Hancu; Ali R Rezai; Andres M Lozano
Journal:  Radiology       Date:  2020-06-23       Impact factor: 11.105

6.  Interactions between thalamic and cortical rhythms during semantic memory recall in human.

Authors:  Scott D Slotnick; Lauren R Moo; Michael A Kraut; Ronald P Lesser; John Hart
Journal:  Proc Natl Acad Sci U S A       Date:  2002-04-23       Impact factor: 11.205

7.  STN DBS of Advanced Parkinson's Disease Experienced in a Specialized Monitoring Unit with a Prospective Protocol.

Authors:  Ji Yeoun Lee; Jung Ho Han; Han Joon Kim; Beom Seok Jeon; Dong Gyu Kim; Sun Ha Paek
Journal:  J Korean Neurosurg Soc       Date:  2008-07-20

8.  Spinal cord stimulator malfunction caused by radiofrequency neuroablation -A case report-.

Authors:  Hye Young Jeon; Jin Woo Shin; Doo Hwan Kim; Jeong Hun Suh; Jeong Gill Leem
Journal:  Korean J Anesthesiol       Date:  2010-12-31

9.  Temperature measurement on neurological pulse generators during MR scans.

Authors:  Wolfgang Kainz; Georg Neubauer; Richard Uberbacher; François Alesch; Dulciana Dias Chan
Journal:  Biomed Eng Online       Date:  2002-09-12       Impact factor: 2.819

10.  Deep brain stimulation for refractory epilepsy.

Authors:  Tomasz Tykocki; Tomasz Mandat; Anna Kornakiewicz; Henryk Koziara; Paweł Nauman
Journal:  Arch Med Sci       Date:  2012-10-08       Impact factor: 3.318

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.