OBJECTIVE: To determine whether pulse oximeter (PO) accuracy and signal quality are affected by level of skin pigmentation.
METHODS: Observational study in a community hospital ED. Consecutive adult patients undergoing arterial blood gas determination were enrolled into the study. Skin pigmentation was determined by comparison with standardized color swatches under controlled lighting; assigned values were used to stratify patients into 3 groups (light, intermediate, and dark) using predetermined criteria. Simultaneous with arterial blood sampling, staff recorded PO reading of O2 saturation using the Nellcor D-25 oximeter. PO values were compared with criterion standard values measured using a 4-wavelength spectrophotometer or co-oximeter. PO signal quality also was recorded. Bias (the mean difference between PO and co-oximeter-measured values of hemoglobin saturation) and precision (the standard deviation of the bias) were calculated. Groups were compared using one-way ANOVA, Bartlett's test for variances, and chi2 test.
RESULTS: O2 saturation data were obtained for 284 patients. Bias values did not differ between the 3 skin pigment groups (p = 0.79). Precision was of borderline significance (p = 0.05), but there was no dose-response relation between skin pigmentation and precision. Study personnel reported suboptimal PO function most often among patients in the dark group (p = 0.003), but this finding was of no clinical significance. PO signal failure was rare (<1% of all patients).
CONCLUSIONS: Although several prior studies suggest the contrary, this study found that skin pigmentation does not affect the bias or precision of pulse oximetry. Furthermore, skin pigmentation has no clinically significant effect on PO signal quality.

Entities:  

Mesh:

Year:  1998        PMID: 9862586     DOI: 10.1111/j.1553-2712.1998.tb02772.x

Source DB:  PubMed          Journal:  Acad Emerg Med        ISSN: 1069-6563            Impact factor:   3.451


  17 in total

1.  An Ethiopian pattern of human adaptation to high-altitude hypoxia.

Authors:  Cynthia M Beall; Michael J Decker; Gary M Brittenham; Irving Kushner; Amha Gebremedhin; Kingman P Strohl
Journal:  Proc Natl Acad Sci U S A       Date:  2002-12-05       Impact factor: 11.205

2.  Effect of newborn screening for critical CHD on healthcare utilisation.

Authors:  Rie Sakai-Bizmark; Hiraku Kumamaru; Eliza J Webber; Dennys Estevez; Laurie A Mena; Emily H Marr; Ruey-Kang R Chang
Journal:  Cardiol Young       Date:  2020-07-02       Impact factor: 1.093

3.  The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia.

Authors:  Elizabeth E Foglia; Robin K Whyte; Aasma Chaudhary; Antonio Mott; Jodi Chen; Kathleen J Propert; Barbara Schmidt
Journal:  J Pediatr       Date:  2016-12-09       Impact factor: 4.406

4.  Wearable Photoplethysmography for Cardiovascular Monitoring.

Authors:  Peter H Charlton; Panicos A Kyriaco; Jonathan Mant; Vaidotas Marozas; Phil Chowienczyk; Jordi Alastruey
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2022-03-11       Impact factor: 10.961

5.  Pulse Oximetry: The Working Principle, Signal Formation, and Applications.

Authors:  Timo Leppänen; Samu Kainulainen; Henri Korkalainen; Saara Sillanmäki; Antti Kulkas; Juha Töyräs; Sami Nikkonen
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 3.650

6.  Influence of Sex, BMI, and Skin Color on the Accuracy of Non-Invasive Cuffless Photoplethysmography-Based Blood Pressure Measurements.

Authors:  Dean Nachman; Arik Eisenkraft; Nir Goldstein; Arik Ben-Ishay; Meir Fons; Roei Merin; Yftach Gepner
Journal:  Front Physiol       Date:  2022-06-29       Impact factor: 4.755

7.  The Big Consequences of Small Discrepancies: Why Racial Differences in Pulse Oximetry Errors Matter.

Authors:  Andre L Holder; An-Kwok Ian Wong
Journal:  Crit Care Med       Date:  2022-02-01       Impact factor: 7.598

8.  A conversion formula for comparing pulse oximeter desaturation rates obtained with different averaging times.

Authors:  Jan Vagedes; Anja Bialkowski; Cornelia Wiechers; Christian F Poets; Klaus Dietz
Journal:  PLoS One       Date:  2014-01-28       Impact factor: 3.240

9.  Racial variations in processes of care for patients with community-acquired pneumonia.

Authors:  Eric M Mortensen; John Cornell; Jeff Whittle
Journal:  BMC Health Serv Res       Date:  2004-08-10       Impact factor: 2.655

10.  A Novel Photoplethysmography Sensor for Vital Signs Monitoring from the Human Trachea.

Authors:  James M May; Justin P Phillips; Tracey Fitchat; Shankar Ramaswamy; Saowarat Snidvongs; Panayiotis A Kyriacou
Journal:  Biosensors (Basel)       Date:  2019-10-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.