Literature DB >> 9844687

MR-guided focused-ultrasound ablation system: determination of precision and accuracy in preparation for clinical trials.

K R Kendell1, T Wu, J P Felmlee, B D Lewis, R L Ehman.   

Abstract

The authors analyzed the accuracy and precision of focal-spot placement with a magnetic resonance-guided, focused-ultrasound system. Average absolute accuracy errors ranged from 0.2 to 1.0 mm, and average absolute individual precision errors ranged from 0.2 to 0.3 mm. To prevent damage to vital structures, single sonications and sonication grids should be placed beyond approximately 2 and 1 mm, respectively.

Mesh:

Year:  1998        PMID: 9844687     DOI: 10.1148/radiology.209.3.9844687

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105