| Literature DB >> 11958655 |
Abstract
In this study, we propose a quality control program for MR-guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the average placement errors were -0.06 mm in the SI direction and -0.04 mm in the LR direction. Temperature elevation was calculated from MR phase difference images and the measured water proton chemical shift (WPCS) temperature coefficient. WPCS temperature calibration for phantoms yielded a temperature coefficient of 0.011 ppm/degree C. Sixteen experiments were conducted using six different phantoms to test the reliability of FUS power delivery. SNR and RF power calculated from phantom images were analyzed and stored at the MR console. A computer program was developed to integrate the system power delivery and the MR image quality control into one automated process. In the clinical trial at our institution, we expect this quality control program to be carried out before each patient treatment. If measured quality control values exceeds or below the preset values, a system service and retest should be conducted before the treatment.Entities:
Mesh:
Year: 2002 PMID: 11958655 PMCID: PMC5724613 DOI: 10.1120/jacmp.v3i2.2584
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1(Color) A diagram of the experiment setup. The water tank is built inside a MR table. The sample in the figure represents the Lucite card or the phantom. When the card was used, it was placed at the focal plane of the focused ultrasound transducer.
Figure 2FUS system placement error measurement in the SI direction. The error value is the difference between the prescribed distance and the actual measured distance between neighboring sonication spots.
Figure 3(Color) WPCS temperature calibration for the phantom.
Figure 4Phase difference images of a phantom during FUS sonication. The bright spot shows the phase change caused by FUS sonication.
Figure 5(Color) The average temperature elevation at FUS focus for 31 tests. The bar shows one standard deviation of the measurement.
Quality control results generated from two tests.
| Transmit gain | Signal | Noise | SNR | Center frequency | Base temperature | Peak temperature | End temperature | Study | Date |
|---|---|---|---|---|---|---|---|---|---|
| 117 | 295 | 6.7 | 43.8 | 9270 | 0.93 | 20.2376 | 9.126 | 2746 | 9/22/01 |
| 116 | 286 | 7.1 | 40.2 | 9420 | 1.199 | 18.879 | 9.071 | 3564 | 10/3/01 |
Tolerance values for the quality control program. The tolerance values are set to be twice the standard deviations for the mean values.
| Transmit gain | Signal | Noise | SNR | Center frequency | Base temperature | Peak temperature | End temperature | |
|---|---|---|---|---|---|---|---|---|
| Mean | 118.00 | 291.24 | 7.77 | 39.47 | 9345.15 | 1.03 | 20.75 | 9.27 |
| Tolerance | 7.14 | 25.49 | 0.92 | 6.50 | 213.49 | 0.67 | 3.27 | 1.74 |