A comparison of the physical imaging properties of Fuji ST-V, ST-VA, and ST-VN computed radiography image plates.

S M Kengyelics1, A G Davies, A R Cowen.   

Abstract

The physical imaging performance of ST-V, ST-VA, and ST-VN computed radiography image plates, read with a Philips AC3 acquisition system, was investigated at 70 kVp with 0.5 mm of copper beam filtration for four different entrance air kerma values: 0.5, 2.6, 26, and 260 micro Gy. Measurements included characteristic response, presampling-modulation transfer function, and Wiener spectrum analysis for (18 x 24) cm2 image plates, sampled at 10 pixels/mm. These results were used to calculate DQE spectral descriptions of system performance. ST-VN image plates exhibited a slightly superior DQE performance for the two lower entrance air kerma values investigated.

Entities:  

Mesh:

Year:  1998        PMID: 9829240     DOI: 10.1118/1.598412

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071