Literature DB >> 9820908

Comparison of the strengths of lumbosacral fixation achieved with techniques using one and two triangulated sacral screws.

J C Leong1, W W Lu, Y Zheng, Q Zhu, S Zhong.   

Abstract

STUDY
DESIGN: The strength of sacral screw fixation achieved with techniques using one sacral screw and two triangulated sacral screws with Chopin block was tested on 10 fresh human sacrum specimens.
OBJECTIVES: To compare the stiffness and the failure strength of single sacral screw fixation and those of two divergent triangulated sacral screw fixation under different loading conditions. SUMMARY OF BACKGROUND DATA: Weak sacral screw fixation is the major cause of failure in instrumented lumbosacral fusion. Although previous studies have evaluated the strength of anteromedially and anterolaterally inserted sacral screws, the strength of two divergent triangulated sacral screws has not been evaluated biomechanically. It is assumed that increasing the number of sacral screws and using a triangulated insertion technique may increase the strength of purchase for lumbosacral fixation.
METHODS: Ten fresh human sacrum specimens were used in this study. Six specimens were from donors less than 30 years of age (younger group), and the other four were from donors more than 60 years of age (aged group). The specimens were assigned randomly to either one side one S1 pedicle screw inserted anteromedially or to the other side with two screws directed anteromedially and anterolaterally. All specimens were tested with an Instron material testing machine (Instron Limited, High Wycombe, England). The average stiffness under compression, tension, and torsion were calculated from the average slopes of the load deformation curves. The failure load was applied for the final testing.
RESULTS: The results showed significant differences in stiffness between fixation with one screw and fixation with two divergent triangulated screws. With one S1 pedicle screw fixation, the average stiffness was 203 N/mm for compression, 147 N/mm for tension, and 2 Nm/degree for torsion. With two-screw fixation the average stiffness increased to 255 N/mm (126%), 185 N/mm (126%) and 2.4 Nm/degree (120%), respectively. The average failure strength under tension loads was found to be 1450 N in the younger specimens and was reduced to 980 N in the aged specimens.
CONCLUSION: This study showed that fixation with two divergent triangulated screws to the sacrum was significantly stronger than one-screw fixation for lumbosacral fusion. The strength of fixation seems to have a negative correlation with aging.

Entities:  

Mesh:

Year:  1998        PMID: 9820908     DOI: 10.1097/00007632-199811010-00008

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  20 in total

1.  A radiological evaluation of the morphometry and safety of S1, S2 and S2-ilium screws in the Asian population using three dimensional computed tomography scan: an analysis of 180 pelvis.

Authors:  Mun Keong Kwan; Amit Jeffry; Chris Yin Wei Chan; Lim Beng Saw
Journal:  Surg Radiol Anat       Date:  2011-12-23       Impact factor: 1.246

2.  [S2-Ala-iliac screws for extended pelvic fixation in longer lumbar instrumentations : Description of a freehand technique].

Authors:  F Lattig; S Weckbach
Journal:  Oper Orthop Traumatol       Date:  2017-06-19       Impact factor: 1.154

3.  Neurovascular risks of sacral screws with bicortical purchase: an anatomical study.

Authors:  Ipek Ergur; Omer Akcali; Amac Kiray; Can Kosay; Hamid Tayefi
Journal:  Eur Spine J       Date:  2007-02-14       Impact factor: 3.134

4.  Pelvic fixation for neuromuscular scoliosis deformity correction.

Authors:  Romain Dayer; Jean Albert Ouellet; Neil Saran
Journal:  Curr Rev Musculoskelet Med       Date:  2012-06

5.  Sacropelvic fixation versus fusion to the sacrum for spondylodesis in multilevel degenerative spine disease.

Authors:  T Finger; S Bayerl; J Onken; M Czabanka; J Woitzik; P Vajkoczy
Journal:  Eur Spine J       Date:  2014-01-22       Impact factor: 3.134

6.  Failure of Posterior Lower Lumbar/Lumbosacral Hemi-Vertebra Resection: An Analysis of Reasons and Revision Strategies.

Authors:  Dun Liu; Benlong Shi; Yang Li; Zhen Liu; Xu Sun; Zezhang Zhu; Yong Qiu
Journal:  Orthop Surg       Date:  2022-06-09       Impact factor: 2.279

7.  Lumbo-sacro-pelvic Fixation Using Iliac Screws for the Complex Lumbo-sacral Fractures.

Authors:  Woo-Tack Rhee; Seung-Hoon You; Yeon-Gyu Jang; Sang-Youl Lee
Journal:  J Korean Neurosurg Soc       Date:  2007-12-20

8.  Reinforcement of lumbosacral instrumentation using S1-pedicle screws combined with S2-alar screws.

Authors:  H Koller; J Zenner; A Hempfing; L Ferraris; O Meier
Journal:  Oper Orthop Traumatol       Date:  2013-06       Impact factor: 1.154

9.  Lumbosacral fixation using the diagonal S2 screw for long fusion in degenerative lumbar deformity: technical note involving 13 cases.

Authors:  Ye-Soo Park; Hong-Sik Kim; Seung-Wook Baek; Sang-Hyun Lee
Journal:  Clin Orthop Surg       Date:  2013-08-20

10.  Biomechanical testing of a unique built-in expandable anterior spinal internal fixation system.

Authors:  Chu-Song Zhou; Yan-Fang Xu; Yu Zhang; Zhong Chen; Hai Lv
Journal:  BMC Musculoskelet Disord       Date:  2014-12-11       Impact factor: 2.362