Literature DB >> 9792574

Pulse oximetry in severe carbon monoxide poisoning.

N B Hampson1.   

Abstract

STUDY
OBJECTIVES: To evaluate the accuracy and quantitate the error of pulse oximetry measurements of arterial oxygenation in patients with severe carbon monoxide (CO) poisoning.
DESIGN: Retrospective review of patient clinical records.
SETTING: Regional referral center for hyperbaric oxygen therapy. PATIENTS: Thirty patients referred for treatment of acute severe CO poisoning who demonstrated carboxyhemoglobin (COHb) levels >25%, with simultaneous determinations of arterial hemoglobin oxygen saturation by pulse oximetry (SpO2) and arterial blood gas (ABG) techniques. MEASUREMENTS AND
RESULTS: COHb levels and measurements of arterial oxygenation from pulse oximetry, ABG analysis, and laboratory CO oximetry were compared. SpO2 did not correlate with COHb levels. SpO2 consistently overestimated the fractional arterial oxygen saturation. The difference between arterial hemoglobin oxygen saturation (SaO2) calculated from ABG analysis and SpO2 increased with increasing COHb level.
CONCLUSIONS: Presently available pulse oximeters overestimate arterial oxygenation in patients with severe CO poisoning. An elevated COHb level falsely elevates the SaO2 measurements from pulse oximetry, usually by an amount less than the COHb level, confirming a prior observation in an animal model. Accurate assessment of arterial oxygen content in patients with CO poisoning can currently be performed only by analysis of arterial blood with a laboratory CO-oximetry.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9792574     DOI: 10.1378/chest.114.4.1036

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  10 in total

Review 1.  Carbon monoxide poisoning.

Authors:  I Blumenthal
Journal:  J R Soc Med       Date:  2001-06       Impact factor: 5.344

Review 2.  [Treatment strategies for acute smoke inhalation injury].

Authors:  D M Maybauer; D L Traber; P Radermacher; D N Herndon; M O Maybauer
Journal:  Anaesthesist       Date:  2006-09       Impact factor: 1.041

3.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

4.  Impact of ethnicity on the accuracy of measurements of oxygen saturations: A retrospective observational cohort study.

Authors:  Mansoor N Bangash; James Hodson; Felicity Evison; Jaimin M Patel; Andrew McD Johnston; Suzy Gallier; Elizabeth Sapey; Dhruv Parekh
Journal:  EClinicalMedicine       Date:  2022-05-06

5.  Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Authors:  John R Feiner; Mark D Rollins; Jeffrey W Sall; Helge Eilers; Paul Au; Philip E Bickler
Journal:  Anesth Analg       Date:  2013-03-11       Impact factor: 5.108

6.  Monitoring in the intensive care.

Authors:  Eric Kipnis; Davinder Ramsingh; Maneesh Bhargava; Erhan Dincer; Maxime Cannesson; Alain Broccard; Benoit Vallet; Karim Bendjelid; Ronan Thibault
Journal:  Crit Care Res Pract       Date:  2012-08-27

7.  Standards of suitability for the management of chronic obstructive respiratory diseases.

Authors:  Claudio M Sanguinetti; Nicolino Ambrosino; Filippo Andò; Fernando De Benedetto; Claudio F Donner; Stefano Nardini; Mario Polverino; Roberto Torchio; Guido Vagheggini; Alberto Visconti
Journal:  Multidiscip Respir Med       Date:  2014-12-18

8.  Detection of hypoxia by near-infrared spectroscopy and pulse oximetry: a comparative study.

Authors:  Amanda Cheung; Lorna Tu; Andrew Macnab; Brian K Kwon; Babak Shadgan
Journal:  J Biomed Opt       Date:  2022-07       Impact factor: 3.758

9.  Ambulatory pulse oximetry monitoring in Japanese COPD outpatients not receiving oxygen therapy.

Authors:  Seigo Minami; Suguru Yamamoto; Yoshitaka Ogata; Takeshi Nakatani; Yoshiko Takeuchi; Masanari Hamaguchi; Taro Koba; Kiyoshi Komuta
Journal:  Multidiscip Respir Med       Date:  2014-04-17

Review 10.  Abnormal fingernail beds following carbon monoxide poisoning: a case report and review of the literature.

Authors:  Mehrangiz Hatami; Frederick Naftolin; Masood A Khatamee
Journal:  J Med Case Rep       Date:  2014-07-29
  10 in total