A S Sarrafzadeh1, K L Kiening, T F Bardt, G H Schneider, A W Unterberg, W R Lanksch.
Abstract
Brain tissue PO2 in severely head injured patients was monitored in parallel with two different PO2-microsensors (Licox and Paratrend). Three different locations of sensor placement were chosen: (1) both catheters into non lesioned tissue (n = 3), (2) both catheters into contusioned tissue (n = 2), and (3) one catheter (Licox) into pericontusional versus one catheter (Paratrend) into non lesioned brain tissue (n = 2). Mean duration of PtiO2-monitoring with both microsensors in parallel was 68.1 hours. Brain tissue PO2 varied when measured in lesioned and nonlesioned tissue. In non lesioned tissue both catheters closely correlated (delta Licox/Paratrend: mean PtiO2 < 5 mm Hg) after 20 hours post insertion. In pericontusional tissue PtiO2 was reduced relative to non lesioned tissue (delta lesioned/non lesioned: mean PtiO2: 10.3 mm Hg). In contusioned brain tissue PtiO2 was always below the "hypoxic threshold" of 10 mm Hg, independent of the type of microsensor used. During a critical reduction in cerebral perfusion pressure (< 60 mm Hg), PtiO2 decreased measured with both microsensors. Elevation of inspired oxygen fraction, normally followed by a rapid increase in tissue PO2, only increased PtiO2 when measured in pericontusional and nonlesioned brain. To recognize critical episodes of hypoxia or ischemia, PtiO2-monitoring of cerebral oxygenation is recommended in nonlesioned brain tissue.Entities:
Mesh:
Substances:
Year: 1998 PMID: 9779180 DOI: 10.1007/978-3-7091-6475-4_54