Literature DB >> 9754181

Characteristics of glaucoma drainage implants during dynamic and steady-state flow conditions.

B A Francis1, A Cortes, J Chen, J A Alvarado.   

Abstract

OBJECTIVE: This study aimed to ascertain whether the Optimed, Krupin, and Ahmed drainage devices function as valves that vary resistance depending on flow conditions to maintain pressure within a desired range. STUDY
DESIGN: Experimental study. INTERVENTION: The three devices and a control cannula were submerged in fluid and perfused with balanced salt solution using a computer-driven apparatus that continuously monitors flow (Q) and pressure (P). In one set of experiments, the flow rates were maintained at 2, 5, 10, 25, or 50 microliters/min until steady-state pressures were achieved. In another set of experiments, the flow rate was increased linearly from 0 to 100 microliters/min over 15 to 20 minutes. MAIN OUTCOME MEASURES: The resistance of each implant was calculated from the first set of experiments by dividing the change in pressure (P) by the change in flow (Q) between successive perfusion rates. Flow-pressure curves were plotted from the experiments in which perfusion rate was increased linearly.
RESULTS: Resistance remained relatively constant for the cannula (0.18-0.24 mmHg/microliter/min), the Krupin (0.09-0.25 mmHg/microliter/min), and the Optimed implants (0.04-0.08) throughout the tested flow rates. For the Ahmed device, conversely, resistance decreased proportionally (2.86-0.05 mmHg/microliter/min) to the increase in flow. When flow rate was increased linearly from 0 to 100 microliters/min, the Optimed and Krupin devices as well as the cannula generated a linear pressure response with a constant slope. The pressure in the two devices increased at a rate of 0.11 mmHg/microliter compared to 0.23 mmHg/microliter/min for the cannula. The flow-pressure curve for the Ahmed implant was distinct with a steep initial pressure rise and an essentially constant pressure of 12 mmHg thereafter.
CONCLUSION: The Optimed and Krupin devices displayed resistance and pressure responses to various flow conditions that were similar to those of a cannula or flow resistor. In these devices, resistance remained relatively stable and pressure increased linearly with flow. The Ahmed device, conversely, functioned as a valve that closely regulated pressure within a desired range by decreasing or increasing resistance as a function of flow.

Entities:  

Mesh:

Year:  1998        PMID: 9754181     DOI: 10.1016/S0161-6420(98)99042-X

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  17 in total

1.  Intraoperative testing of opening and closing pressure predicts risk of low intraocular pressure after Ahmed glaucoma valve implantation.

Authors:  F Bochmann; A Kipfer; J Tarantino; C Kaufmann; L Bachmann; M Thiel
Journal:  Eye (Lond)       Date:  2014-07-25       Impact factor: 3.775

2.  Intraocular pressure control of a novel glaucoma drainage device - in vitro and in vivo studies.

Authors:  Li-Jun Cui; Di-Chen Li; Jian Liu; Lei Zhang; Yao Xing
Journal:  Int J Ophthalmol       Date:  2017-09-18       Impact factor: 1.779

3.  Control and optimisation of fluid flow in glaucoma drainage device surgery.

Authors:  Kin Sheng Lim
Journal:  Eye (Lond)       Date:  2018-01-19       Impact factor: 3.775

4.  The outcome of the Ahmad Glaucoma Valve implantation for refractory glaucoma in Kuwait.

Authors:  Adel M Aljazzaf; Sidky M A Abdelmoaty; Abdelmutalib H Behbehani; Ahmad A Abdulmuez; Hadeel A Aljazzaf
Journal:  Saudi J Ophthalmol       Date:  2012-10-27

5.  Laser Activated Flow Regulator for Glaucoma Drainage Devices.

Authors:  Jeffrey L Olson; Raul Velez-Montoya; Ramanath Bhandari
Journal:  Transl Vis Sci Technol       Date:  2014-11-03       Impact factor: 3.283

6.  Clinical outcomes of Ahmed glaucoma valve implantation using tube ligation and removable external stents.

Authors:  Jong Joo Lee; Ki Ho Park; Dong Myung Kim; Tae Woo Kim
Journal:  Korean J Ophthalmol       Date:  2009-06-09

Review 7.  The Ahmed shunt versus the Baerveldt shunt for refractory glaucoma: a meta-analysis.

Authors:  Shiming Wang; Xiaoming Gao; Nana Qian
Journal:  BMC Ophthalmol       Date:  2016-06-08       Impact factor: 2.209

8.  Performance of and Pressure Elevation Formed by Small-diameter Microtubes Used in Constant-flow Sets.

Authors:  Sae Woon Sohn; Myounggyu D Noh; Jong-Hyun Lee; Kyoung Nam Kim; Chang-Sik Kim; Byung Heon Ahn
Journal:  Korean J Ophthalmol       Date:  2016-05-18

9.  Current surgical options for the management of pediatric glaucoma.

Authors:  Jose Morales; Sami Al Shahwan; Sami Al Odhayb; Ibrahim Al Jadaan; Deepak P Edward
Journal:  J Ophthalmol       Date:  2013-04-24       Impact factor: 1.909

10.  Results of a New Technique for Implantation of Nonrestrictive Glaucoma Devices.

Authors:  Gabriel Enrique Ortiz Arismendi; Cristina Del Pilar Peña Valderrama; Oscar Albis-Donado
Journal:  J Curr Glaucoma Pract       Date:  2013-09-06
View more