Literature DB >> 29350692

Control and optimisation of fluid flow in glaucoma drainage device surgery.

Kin Sheng Lim1.   

Abstract

Avoidance of hypotony in the early post-operative phase following glaucoma drainage device surgery is paramount, if serious complications are to be avoided. Current strategies of preventing early hypotony are either a valve mechanism, as those found in Ahmed Glaucoma Valve or flow restrictor as used in Baerveldt or Molteno implants. The Ahmed glaucoma valve has a pre-tensioned valve which opens and closes at the pre-determined level of intra-ocular pressure. Although the opening and closing pressure can be slightly inconsistent, coinciding with the change of material used in the plate construction, the reliability appears to have improved in recent years. Flow restrictor method, such as vicryl tie and Supramid suture stenting, used in non-valved implants utilised the Poissuille's equation in predicting the pressure gradient across the tube and thereby avoiding the risk of hypotony. Newer generations of glaucoma drainage devices such as Xen, Microshunt, and eyeWatch all use the same principle. However, most of these do not address the issue of unnecessary residual flow resistance once the bleb resistance has formed around the drainage tube or plate except for the EyeWatch device.

Entities:  

Mesh:

Year:  2018        PMID: 29350692      PMCID: PMC5811747          DOI: 10.1038/eye.2017.316

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  20 in total

1.  Application of Poiseuille's law to aqueous outflow.

Authors:  W K McEWEN
Journal:  AMA Arch Ophthalmol       Date:  1958-08

2.  In vitro flow testing of glaucoma drainage devices.

Authors:  J M Porter; C H Krawczyk; R F Carey
Journal:  Ophthalmology       Date:  1997-10       Impact factor: 12.079

3.  Comparative human aqueous dynamics study between black and white subjects with glaucoma.

Authors:  Laura Beltran-Agullo; Pouya Alaghband; Safina Rashid; Josee Gosselin; Adanna Obi; Rahat Husain; Kin Sheng Lim
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-12-09       Impact factor: 4.799

4.  In vitro flow properties of glaucoma implant devices.

Authors:  D L Eisenberg; E Y Koo; G Hafner; J S Schuman
Journal:  Ophthalmic Surg Lasers       Date:  1999 Sep-Oct

5.  The vicryl tie technique for inserting a draining implant in the treatment of secondary glaucoma.

Authors:  A C Molteno; P J Polkinghorne; J A Bowbyes
Journal:  Aust N Z J Ophthalmol       Date:  1986-11

6.  New implant for drainage in glaucoma. Animal trial.

Authors:  A C Molteno
Journal:  Br J Ophthalmol       Date:  1969-03       Impact factor: 4.638

7.  Experimental flow studies in glaucoma drainage device development.

Authors: 
Journal:  Br J Ophthalmol       Date:  2001-10       Impact factor: 4.638

8.  Initial clinical experience with the Ahmed Glaucoma Valve implant.

Authors:  A L Coleman; R Hill; M R Wilson; N Choplin; R Kotas-Neumann; M Tam; J Bacharach; W C Panek
Journal:  Am J Ophthalmol       Date:  1995-07       Impact factor: 5.258

9.  Delayed nonexpulsive suprachoroidal hemorrhage.

Authors:  M G Gressel; R K Parrish; D K Heuer
Journal:  Arch Ophthalmol       Date:  1984-12

Review 10.  Flow of aqueous humor in humans [The Friedenwald Lecture].

Authors:  R F Brubaker
Journal:  Invest Ophthalmol Vis Sci       Date:  1991-12       Impact factor: 4.799

View more
  5 in total

1.  Comparison of Two Tube-Modification Techniques in Baerveldt Glaucoma Implantation in Refractory Glaucoma.

Authors:  Purit Petpiroon; Thanyathorn Vijittrakarnrung; Wasu Supakontanasan; Apichat Tantraworasin; Yanin Suwan
Journal:  Clin Ophthalmol       Date:  2020-06-19

2.  Evaluation of the Ultrastructural and In Vitro Flow Properties of the PRESERFLO MicroShunt.

Authors:  Marta Ibarz Barberá; Jose Luis Hernández-Verdejo; Jean Bragard; Javier Burguete; Laura Morales Fernández; Pedro Tañá Rivero; Rosario Gómez de Liaño; Miguel A Teus
Journal:  Transl Vis Sci Technol       Date:  2021-11-01       Impact factor: 3.283

3.  A model for designing intraocular pressure-regulating glaucoma implants.

Authors:  Inês C F Pereira; Hans M Wyss; Leonard Pinchuk; Henny J M Beckers; Jaap M J den Toonder
Journal:  PLoS One       Date:  2022-09-02       Impact factor: 3.752

4.  Cataract and glaucoma combined surgery: XEN® gel stent versus nonpenetrating deep sclerectomy, a pilot study.

Authors:  Vincent Theillac; Esther Blumen-Ohana; Jad Akesbi; Pascale Hamard; Alexandre Sellam; Emmanuelle Brasnu; Christophe Baudouin; Antoine Labbe; Jean-Philippe Nordmann
Journal:  BMC Ophthalmol       Date:  2020-06-16       Impact factor: 2.209

Review 5.  Conventional glaucoma implants and the new MIGS devices: a comprehensive review of current options and future directions.

Authors:  Inês C F Pereira; Rosanne van de Wijdeven; Hans M Wyss; Henny J M Beckers; Jaap M J den Toonder
Journal:  Eye (Lond)       Date:  2021-06-14       Impact factor: 3.775

  5 in total