Accurate determination of effective lens position and lens-capsule distance with 4 intraocular lenses.

O Findl1, W Drexler, R Menapace, B Bobr, S Bittermann, C Vass, G Rainer, C K Hitzenberger, A F Fercher.   

Abstract

PURPOSE: To measure effective lens position (ELP) of 4 intraocular lenses (IOLs) using high precision and high resolution dual-beam partial coherence interferometry (PCI) and to assess the tendency of these IOLs to produce a lens-capsule distance (LCD), a possible risk factor for posterior capsule opacification.
SETTING: Department of Ophthalmology, Vienna General Hospital; Institute of Medical Physics, University of Vienna, Austria.
METHODS: In a retrospective study, PCI was used to measure ELP and LCD in 139 pseudophakic eyes of 110 patients with 4 IOLs: acrylic 3-piece IOL (AcrySof MA60BM); silicone 3-piece IOL without a capsular tension ring (PhacoFlex SI30) and with a capsular tension ring (PhacoFlex SI30 and Morcher Type 14); silicone plate-haptic IOL (Staar AA4203VF); and a hydrogel plate-haptic IOL (logel 1103).
RESULTS: The ELP and LCD were determined with a precision of approximately 3 to 4 microns. An LCD was detected in 21% eyes with the AcrySof, 20% of eyes with the SI30 without a capsular tension ring, 10% of eyes with a capsular tension ring, 21% of eyes with the Staar, and 17% of eyes with the logel. The LCDs detected by PCI, but not by slitlamp examination, were significantly smaller than those detected by both.
CONCLUSION: The amount of LCD detected by PCI was approximately the same with all IOL types (approximately 20%) except the PhacoFlex SI30 with a capsular tension ring (10%).

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9719969     DOI: 10.1016/s0886-3350(98)80103-x

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  [Three-dimensional ultrasound biomicroscopic examinations for haptic differentiation of potentially accommodative intraocular lenses].

Authors:  O Stachs; H Schneider; J Stave; R Beck; R F Guthoff
Journal:  Ophthalmologe       Date:  2005-03       Impact factor: 1.059

2.  A study to evaluate whether CTR increases refractive unpredictability between predicted and actual IOL position.

Authors:  V K Baranwal; Santosh Kumar; Avinash Mishra; Ajay K Dutta
Journal:  Med J Armed Forces India       Date:  2012-10-31

3.  Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode.

Authors:  B Meinhardt; O Stachs; J Stave; R Beck; R Guthoff
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-09-15       Impact factor: 3.117

4.  Effect of the capsular tension ring on refractive outcome after phacoemulsification.

Authors:  Gültekin Irgat Saadet; Özcura Fatih
Journal:  Rom J Ophthalmol       Date:  2021 Jan-Mar

5.  Effect of capsular tension ring implantation on capsular stability after phacoemulsification in patients with weak zonules: a randomized controlled trial. CTR implantation in cataract patients with weak zonules.

Authors:  Shangfei Yang; Hui Jiang; Kailai Nie; Liwen Feng; Wei Fan
Journal:  BMC Ophthalmol       Date:  2021-01-07       Impact factor: 2.209

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.