Reliability of epidural pressure measurement in clinical practice: behavior of three modern sensors during simultaneous ipsilateral intraventricular or intraparenchymal pressure measurement.

A Raabe1, R Totzauer, O Meyer, R Stöckel, D Hohrein, J Schöche.   

Abstract

OBJECTIVE: In recent years, new devices for epidural pressure (EDP) measurement have been developed, with claims of improved accuracy. However, there have been no new data from simultaneous pressure measurements to reverse the skepticism regarding this technique.
METHODS: The reliability of EDP recording was investigated in 26 patients with severe head injuries, during simultaneous measurement of ipsilateral intraventricular pressure (IVP) or intraparenchymal pressure (IPP).
RESULTS: Periods of simultaneous measurement ranged from 3.3 to 168 hours. Technical failure occurred in five cases. Clinically significant erroneous pressure values (defined as differences of >10 mm Hg between EDP and IVP or IPP, lasting for >10 min) occurred in 7 of 21 measurements (33%). In three of these seven cases (43%) the difference between the EDP and the IVP or IPP remained nearly stable throughout the period of observation, whereas in four cases (57%) significant drifting over time was observed. Analysis of the transfer function calculated for the first to sixth harmonics between the EDP and the IVP or IPP showed no significant differences between the pulse pressure waveforms, even when significant drifting of more than 20 mm Hg occurred.
CONCLUSION: The reliability of EDP measurement has not improved, compared with 10 years ago. Using this technique, one must be aware of the possibility of overestimating intracranial pressure by more than 10 mm Hg and the possibility of significant drifting of EDP over time, which argues against the use of this method even for trend analysis.

Entities:  

Mesh:

Year:  1998        PMID: 9696083     DOI: 10.1097/00006123-199808000-00073

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  4 in total

1.  Intracranial pressure complicating severe traumatic brain injury in children: monitoring and management.

Authors:  Kevin P Morris; Robert J Forsyth; Roger C Parslow; Robert C Tasker; Carol A Hawley
Journal:  Intensive Care Med       Date:  2006-07-28       Impact factor: 17.440

Review 2.  Neuromonitoring in the intensive care unit. I. Intracranial pressure and cerebral blood flow monitoring.

Authors:  Anuj Bhatia; Arun Kumar Gupta
Journal:  Intensive Care Med       Date:  2007-05-24       Impact factor: 17.440

Review 3.  Imaging Guidance for Therapeutic Delivery: The Dawn of Neuroenergetics.

Authors:  Vilakshan Alambyan; Jonathan Pace; Persen Sukpornchairak; Xin Yu; Hamza Alnimir; Ryan Tatton; Gautham Chitturu; Anisha Yarlagadda; Ciro Ramos-Estebanez
Journal:  Neurotherapeutics       Date:  2020-04       Impact factor: 7.620

Review 4.  Accuracy of intracranial pressure monitoring: systematic review and meta-analysis.

Authors:  Lucia Zacchetti; Sandra Magnoni; Federica Di Corte; Elisa R Zanier; Nino Stocchetti
Journal:  Crit Care       Date:  2015-12-02       Impact factor: 9.097