Analysis of low-dose digital lateral cephalometric radiographs.

E B Näslund1, M Kruger, A Petersson, K Hansen.   

Abstract

OBJECTIVE: To examine the effects of the dose reduction obtained with photostimulable phosphor computed radiography (CR) on the identification of cephalometric landmarks.
METHODS: Dose reductions of 50% and 75% respectively were obtained by placing three storage phosphor image plates in the same cassette. Lateral cephalograms were exposed on ten patients. Seven observers identified 14 anatomical landmarks on the images.
RESULTS: The location of anatomical landmarks on the lateral cephalograms was not effected by the decreased dose (P > 0.05). Interobserver variation was the major influence.
CONCLUSION: A dose reduction of 75% does not effect the localization of anatomical landmarks in lateral cephalograms obtained with CR.

Entities:  

Mesh:

Year:  1998        PMID: 9693525     DOI: 10.1038/sj/dmfr/4600333

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  3 in total

1.  Prospective study on the reproducibility of cephalometric landmarks on conventional and digital lateral headfilms.

Authors:  K Hagemann; D Vollmer; T Niegel; U Ehmer; I Reuter
Journal:  J Orofac Orthop       Date:  2000       Impact factor: 1.938

2.  Reliability Assessment of Orthodontic Apps for Cephalometrics.

Authors:  Sertaç Aksakallı; Hilal Yılancı; Erhan Görükmez; Sabri İlhan Ramoğlu
Journal:  Turk J Orthod       Date:  2016-12-01

3.  Comparison between cephalometric measurements using digital manual and web-based artificial intelligence cephalometric tracing software.

Authors:  Gökhan Çoban; Taner Öztürk; Nizami Hashimli; Ahmet Yağci
Journal:  Dental Press J Orthod       Date:  2022-08-15
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.