Literature DB >> 9659595

Magnetic resonance diagnosis of the anterior labrum and capsule. Effect of field strength on efficacy.

K H Allmann1, O Walter, J Laubenberger, M Uhl, C H Buitrago-Tellez, N Biebow, M Langer.   

Abstract

RATIONALE AND
OBJECTIVES: The authors compare the performance of high-field and low-field magnetic resonance (MR) systems in the evaluation of shoulder instability, 35 patients were examined at field strengths of 1.0 T and of 0.2 T.
METHODS: Surface coils were used in both systems. Because arthroscopy was used as gold standard, a preselected patient-population was obtained for the study.
RESULTS: The sensitivity/specificity/accuracy of MR images acquired at 1.0 T for labrum pathology were 91%/67%/91% and 70%/80%/71% for the capsular complex. Compared with the above, the sensitivity/specificity/accuracy for 0.2 T MR images revealed 91%/67%/91% for the labrum pathology and 63%/80%/66% for the capsular complex respectively. In the evaluation of capsular lesions a comparison between the 0.2 T MR system and the 1.0 T system indicated a higher sensitivity and accuracy for the high-field images. Concerning labral lesions, the sensitivity and accuracy of the 0.2 T MR imager and the 1.0 T imager were comparable.
CONCLUSIONS: Given differences in imaging protocols, imaging at 0.2 T does not adversely affect the assessment of shoulder instability when compared with imaging at 1.0 T. These preliminary results warrant more extensive clinical comparison of results obtained at different magnetic field strengths.

Entities:  

Mesh:

Year:  1998        PMID: 9659595     DOI: 10.1097/00004424-199807000-00007

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  4 in total

1.  MR arthrography of the shoulder: comparison of low-field (0.2 T) vs high-field (1.5 T) imaging.

Authors:  R Loew; K F Kreitner; M Runkel; J Zoellner; M Thelen
Journal:  Eur Radiol       Date:  2000       Impact factor: 5.315

2.  Non-cryogenic anatomical imaging in ultra-low field regime: hand MRI demonstration.

Authors:  I Savukov; T Karaulanov; A Castro; P Volegov; A Matlashov; A Urbatis; J Gomez; M Espy
Journal:  J Magn Reson       Date:  2011-06-01       Impact factor: 2.229

3.  Non-cryogenic ultra-low field MRI of wrist-forearm area.

Authors:  I Savukov; T Karaulanov; C J V Wurden; L Schultz
Journal:  J Magn Reson       Date:  2013-06-07       Impact factor: 2.229

4.  Analysis of Low-Field MRI Scanners for Evaluation of Shoulder Pathology Based on Arthroscopy.

Authors:  Christopher S Lee; Shane M Davis; Claire McGroder; Shalen Kouk; Ryan M Sung; William B Stetson; Scott E Powell
Journal:  Orthop J Sports Med       Date:  2014-07-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.