Adaptive canceling of physiological tremor for improved precision in microsurgery.

C N Riviere1, R S Rader, N V Thakor.   

Abstract

Physiological hand tremor impedes microsurgery. We present both a novel adaptive algorithm for tremor estimation and a new technique for active real-time canceling of physiological tremor. Tremor is modeled online using the weighted-frequency Fourier linear combiner (WFLC). This adaptive algorithm models tremor as a modulating sinusoid, and tracks its frequency, amplitude, and phase. Piezoelectric actuators move the surgical instrument tip in opposition to the motion of tremor, effectively subtracting the tremor from the total motion. We demonstrate the technique in one dimension using a cantilever apparatus as a benchtop simulation of the surgical instrument. Actual hand motion, prerecorded during simulated surgery, is used as input. In 25 tests, WFLC tremor compensation reduces the rms tip motion in the 6-16 Hz tremor band by 67%, and reduces the rms error with respect to an a posteriori estimate of voluntary motion by 30%. The technique can be implemented in a hand-held microsurgical instrument.

Entities:  

Mesh:

Year:  1998        PMID: 9644892     DOI: 10.1109/10.686791

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  20 in total

1.  Augmented visual feedback increases finger tremor during postural pointing.

Authors:  J Keogh; S Morrison; R Barrett
Journal:  Exp Brain Res       Date:  2004-07-30       Impact factor: 1.972

2.  Robotic Motion Compensation for Beating Heart Intracardiac Surgery.

Authors:  Shelten G Yuen; Daniel T Kettler; Paul M Novotny; Richard D Plowes; Robert D Howe
Journal:  Int J Rob Res       Date:  2009-10-01       Impact factor: 4.703

3.  Spinal cord integrity monitoring by adaptive coherence measurement.

Authors:  D L Sherman; V Wuyyuru; M Jason Brooke; H X Zhang; J P Sepkuty; N V Thakor; A Natarajan; A H All
Journal:  J Neurosci Methods       Date:  2010-08-06       Impact factor: 2.390

4.  Vision-based proximity detection in retinal surgery.

Authors:  R Richa; M Balicki; R Sznitman; E Meisner; R Taylor; G Hager
Journal:  IEEE Trans Biomed Eng       Date:  2012-06-05       Impact factor: 4.538

Review 5.  Joystick control for powered mobility: current state of technology and future directions.

Authors:  Brad E Dicianno; Rory A Cooper; John Coltellaro
Journal:  Phys Med Rehabil Clin N Am       Date:  2010-02       Impact factor: 1.784

6.  Extraction of user's navigation commands from upper body force interaction in walker assisted gait.

Authors:  Anselmo Frizera Neto; Juan A Gallego; Eduardo Rocon; José L Pons; Ramón Ceres
Journal:  Biomed Eng Online       Date:  2010-08-05       Impact factor: 2.819

7.  Estimating Displacement of Periodic Motion With Inertial Sensors.

Authors:  U-Xuan Tan; Kalyana C Veluvolu; Win Tun Latt; Cheng Yap Shee; Cameron N Riviere; Wei Tech Ang
Journal:  IEEE Sens J       Date:  2009-08-01       Impact factor: 3.301

8.  Predictive Filtering in Motion Compensation with Steerable Cardiac Catheters.

Authors:  Paul M Loschak; Alperen Degirmenci; Robert D Howe
Journal:  IEEE Int Conf Robot Autom       Date:  2017-07-24

9.  Use of weighted Fourier linear combiner filters to estimate lower trunk 3D orientation from gyroscope sensors data.

Authors:  Vincent Bonnet; Claudia Mazzà; John McCamley; Aurelio Cappozzo
Journal:  J Neuroeng Rehabil       Date:  2013-03-11       Impact factor: 4.262

10.  A neuroprosthesis for tremor management through the control of muscle co-contraction.

Authors:  Juan Álvaro Gallego; Eduardo Rocon; Juan Manuel Belda-Lois; José Luis Pons
Journal:  J Neuroeng Rehabil       Date:  2013-04-15       Impact factor: 4.262

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.