Abstract

We designed a novel guide catheter specifically for interventions to the left coronary artery via a right upper limb approach. The catheter has a novel first loop design which utilizes the angle between the right subclavian and innominate arteries for support. The first loop introduces the catheter into the correct position and generates powerful and coaxial back-up power. We report successful implantation of Palmaz-Schatz stents in five cases using this 6 Fr. catheter.

Entities:  

Mesh:

Year:  1998        PMID: 9637453     DOI: 10.1002/(sici)1097-0304(199806)44:2<244::aid-ccd24>3.0.co;2-l

Source DB:  PubMed          Journal:  Cathet Cardiovasc Diagn        ISSN: 0098-6569


  4 in total

1.  Complete fracture of an Ikari guiding catheter in the axillary artery during transradial coronary intervention.

Authors:  Yian-Ping Lee; Huay-Cheem Tan; Chi-Hang Lee
Journal:  Int J Angiol       Date:  2008

Review 2.  Coronary angiography in the angioplasty era: projections with a meaning.

Authors:  Carlo Di Mario; Nilesh Sutaria
Journal:  Heart       Date:  2005-07       Impact factor: 5.994

3.  Novel diagnostic catheter specifically designed for both coronary arteries via the right transradial approach. A prospective, randomized trial of Tiger II vs. Judkins catheters.

Authors:  Seong-Man Kim; Dae-Kyeong Kim; Doo-Il Kim; Dong-Soo Kim; Seung-Jae Joo; Jae-Woo Lee
Journal:  Int J Cardiovasc Imaging       Date:  2005-11-22       Impact factor: 2.357

4.  Efficacy of a one-catheter concept for transradial coronary angiography.

Authors:  Christoph Langer; Julia Riehle; Helge Wuttig; Stephanie Dürrwald; Helmut Lange; Alexander Samol; Norbert Frey; Marcus Wiemer
Journal:  PLoS One       Date:  2018-01-02       Impact factor: 3.240