Complete fracture of an Ikari guiding catheter in the axillary artery during transradial coronary intervention.

Yian-Ping Lee1, Huay-Cheem Tan, Chi-Hang Lee.   

Abstract

Complete fracture of a guiding catheter during percutaneous coronary intervention is an exceedingly rare but potentially serious complication. The present case report describes the complete fracture of an Ikari guiding catheter and the various manoeuvres used to retrieve the broken catheter fragment. The anatomy and limited space in the radial and axillary arteries made it possible for the catheter to twist on itself during a difficult cannulation, resulting in its complete fracture and separation. The present case highlights the need for caution in the manipulation of an Ikari guiding catheter by the transradial approach.

Entities:  

Keywords:  Axillary artery; Coronary; Fracture; Ikari guiding catheter; Intervention; Transradial

Year:  2008        PMID: 22477371      PMCID: PMC2728372          DOI: 10.1055/s-0031-1278279

Source DB:  PubMed          Journal:  Int J Angiol        ISSN: 1061-1711


  5 in total

1.  Initial characterization of Ikari Guide catheter for transradial coronary intervention.

Authors:  Yuji Ikari; Hiroyoshi Nakajima; Raisuke Iijima; Jiro Aoki; Kengo Tanabe; Tomohiro Nakayama; Akiyoshi Miyazawa; Mitsuharu Hatori; Hiroyuki Kyouno; Shuzou Tanimoto; Eisuke Amiya; Gaku Nakazawa; Yoshinobu Onuma; Kazuhiro Hara
Journal:  J Invasive Cardiol       Date:  2004-02       Impact factor: 2.022

2.  The physics of guiding catheters for the left coronary artery in transfemoral and transradial interventions.

Authors:  Yuji Ikari; Masakazu Nagaoka; Jae-Young Kim; Yoshihiro Morino; Teruhisa Tanabe
Journal:  J Invasive Cardiol       Date:  2005-12       Impact factor: 2.022

3.  Novel guide catheter for left coronary intervention via a right upper limb approach.

Authors:  Y Ikari; M Ochiai; M Hangaishi; M Ohno; J Taguchi; K Hara; T Isshiki; T Tamura; T Yamaguchi
Journal:  Cathet Cardiovasc Diagn       Date:  1998-06

4.  A randomized comparison of percutaneous transluminal coronary angioplasty by the radial, brachial and femoral approaches: the access study.

Authors:  F Kiemeneij; G J Laarman; D Odekerken; T Slagboom; R van der Wieken
Journal:  J Am Coll Cardiol       Date:  1997-05       Impact factor: 24.094

5.  A randomized study comparing same-day home discharge and abciximab bolus only to overnight hospitalization and abciximab bolus and infusion after transradial coronary stent implantation.

Authors:  Olivier F Bertrand; Robert De Larochellière; Josep Rodés-Cabau; Guy Proulx; Onil Gleeton; Can Manh Nguyen; Jean-Pierre Déry; Gérald Barbeau; Bernard Noël; Eric Larose; Paul Poirier; Louis Roy
Journal:  Circulation       Date:  2006-12-04       Impact factor: 29.690

  5 in total
  1 in total

1.  Foreign body associated non-ST elevation myocardial infarction: a case report.

Authors:  Pooyan Dehghani; Mahdi Sajedi Khanian
Journal:  J Cardiovasc Thorac Res       Date:  2017-12-20
  1 in total