[Dosage measurements in spiral CT examinations of the head and neck region].

M Cohnen1, B Cohnen, K Ewen, G Teubert, U Mödder.   

Abstract

PURPOSE: In view of the increasing frequency of computed tomographic examinations the organ doses of the thyroid gland and the eye lenses in spiral CT examinations of the head and neck should be determined.
MATERIAL AND METHODS: Standard CT-examinations of the neck, the paranasal sinuses in axial and coronal sections, and the mandibula were simulated by help of an Alderson-Rando phantom. Dose measurements were performed as well in 28 patients. The examinations were carried out in spiral mode ("Somatom AR.SP", "Picker PQ 2000", and "Toshiba Xpress SX") and conventional single slice technique ("Somatom Plus"). Three LiF-TLD were placed on each eyelid and three on the thyroid.
RESULTS: With direct exposure to the x-ray-beam the organ doses did not exceed 64 mGy. When exposed by scatter radiation only, doses may be reduced by a factor of up to 10. The measurements in 28 patients showed similar results.
CONCLUSIONS: The organ doses in spiral CT correspond with data known from conventional CT.

Entities:  

Mesh:

Year:  1998        PMID: 9617364     DOI: 10.1055/s-2007-1015164

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  2 in total

1.  CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction.

Authors:  M Cohnen; H Fischer; J Hamacher; E Lins; R Kötter; U Mödder
Journal:  AJNR Am J Neuroradiol       Date:  2000-10       Impact factor: 3.825

2.  Efficacy of breast shielding during CT of the head.

Authors:  Z Brnić; B Vekić; A Hebrang; P Anić
Journal:  Eur Radiol       Date:  2003-06-25       Impact factor: 5.315

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.