Calibration of a mosfet detection system for 6-MV in vivo dosimetry.

P Scalchi1, P Francescon.   

Abstract

PURPOSE: Metal oxide semiconductor field-effect transistor (MOSFET) detectors were calibrated to perform in vivo dosimetry during 6-MV treatments, both in normal setup and total body irradiation (TBI) conditions. METHODS AND MATERIALS: MOSFET water-equivalent depth, dependence of the calibration factors (CFs) on the field sizes, MOSFET orientation, bias supply, accumulated dose, incidence angle, temperature, and spoiler-skin distance in TBI setup were investigated. MOSFET reproducibility was verified. The correlation between the water-equivalent midplane depth and the ratio of the exit MOSFET readout divided by the entrance MOSFET readout was studied. MOSFET midplane dosimetry in TBI setup was compared with thermoluminescent dosimetry in an anthropomorphic phantom. By using ionization chamber measurements, the TBI midplane dosimetry was also verified in the presence of cork as a lung substitute.
RESULTS: The water-equivalent depth of the MOSFET is about 0.8 mm or 1.8 mm, depending on which sensor side faces the beam. The field size also affects this quantity; Monte Carlo simulations allow driving this behavior by changes in the contaminating electron mean energy. The CFs vary linearly as a function of the square field side, for fields ranging from 5 x 5 to 30 x 30 cm2. In TBI setup, varying the spoiler-skin distance between 5 mm and 10 cm affects the CFs within 5%. The MOSFET reproducibility is about 3% (2 SD) for the doses normally delivered to the patients. The effect of the accumulated dose on the sensor response is negligible. For beam incidence ranging from 0 degrees to 90 degrees, the MOSFET response varies within 7%. No monotonic correlation between the sensor response and the temperature is apparent. Good correlation between the water-equivalent midplane depth and the ratio of the exit MOSFET readout divided by the entrance MOSFET readout was found (the correlation coefficient is about 1). The MOSFET midplane dosimetry relevant to the anthropomorphic phantom irradiation is in agreement with TLD dosimetry within 5%. Ionization chamber and MOSFET midplane dosimetry in inhomogeneous phantoms are in agreement within 2%.
CONCLUSION: MOSFET characteristics are suitable for the in vivo dosimetry relevant to 6-MV treatments, both in normal and TBI setup. The TBI midplane dosimetry using MOSFETs is valid also in the presence of the lung, which is the most critical organ, and allows verifying that calculation of the lung attenuator thicknesses based only on the density is not correct. Our MOSFET dosimetry system can be used also to determine the surface dose by using the water-equivalent depth and extrapolation methods. This procedure depends on the field size used.

Entities:  

Mesh:

Year:  1998        PMID: 9531385     DOI: 10.1016/s0360-3016(97)00894-8

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  9 in total

1.  Practical issues regarding angular and energy response in in vivo intraoperative electron radiotherapy dosimetry.

Authors:  Juan López-Tarjuelo; Ana Bouché-Babiloni; Virginia Morillo-Macías; Agustín Santos-Serra; Carlos Ferrer-Albiach
Journal:  Rep Pract Oncol Radiother       Date:  2016-10-19

2.  In vivo dosimetry in intraoperative electron radiotherapy: microMOSFETs, radiochromic films and a general-purpose linac.

Authors:  Juan López-Tarjuelo; Ana Bouché-Babiloni; Virginia Morillo-Macías; Noelia de Marco-Blancas; Agustín Santos-Serra; Juan David Quirós-Higueras; Carlos Ferrer-Albiach
Journal:  Strahlenther Onkol       Date:  2014-06-26       Impact factor: 3.621

3.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

4.  On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy.

Authors:  Fasihah Hanum Yusof; Ngie Min Ung; Jeannie Hsiu Ding Wong; Wei Loong Jong; Vannyat Ath; Vincent Chee Ee Phua; Siew Ping Heng; Kwan Hoong Ng
Journal:  PLoS One       Date:  2015-06-08       Impact factor: 3.240

5.  Response of Nanodot Optically Stimulated Luminescence Dosimeters to Therapeutic Electron Beams.

Authors:  Y Retna Ponmalar; Ravikumar Manickam; S Sathiyan; K M Ganesh; R Arun; Henry Finlay Godson
Journal:  J Med Phys       Date:  2017 Jan-Mar

6.  Light-Dependent Resistors as Dosimetric Sensors in Radiotherapy.

Authors:  Juan Román-Raya; Isidoro Ruiz-García; Pablo Escobedo; Alberto J Palma; Damián Guirado; Miguel A Carvajal
Journal:  Sensors (Basel)       Date:  2020-03-11       Impact factor: 3.576

7.  Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams.

Authors:  A Sathish Kumar; S D Sharma; B Paul Ravindran
Journal:  J Med Phys       Date:  2014-07

8.  Characterization of MOSkin detector for in vivo skin dose measurement during megavoltage radiotherapy.

Authors:  Wei Loong Jong; Jeannie Hsiu Ding Wong; Ngie Min Ung; Kwan Hoong Ng; Gwo Fuang Ho; Dean L Cutajar; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

9.  In vivo prostate IMRT dosimetry with MOSFET detectors using brass buildup caps.

Authors:  Raj Varadhan; John Miller; Brenden Garrity; Michael Weber
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.