| Literature DB >> 17533354 |
Raj Varadhan1, John Miller, Brenden Garrity, Michael Weber.
Abstract
The feasibility of using dual bias metal oxide semiconductor field effect transistors (MOSFET) detectors with the new hemispherical brass build up cap for in vivo dose measurements in Prostate IMRT treatments was investigated and achieved. A general formalism with various correction factors taken into account to predict Dmax entrance skin dose for the IMRT fields with MOSFETS was developed and compared against predicted dose from the treatment planning system. We achieved an overall accuracy better than +/- 5 % on all measured fields for both 6 MV and 10 MV beams when compared to predicted doses from Philips Pinnacle3 and CMS XiO treatment planning systems respectively. We also estimate the total uncertainty in estimation of MOSFET dose in the high sensitivity mode for IMRT therapy to be 4.6%.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17533354 PMCID: PMC5722390 DOI: 10.1120/jacmp.v7i4.2278
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1MOSFET AutoSense System
Figure 2The brass buildup cap has a radius of 0.635 cm and a special groove insert for MOSFET placement, which improves accuracy of placement.
Correction factors (CR) for MOSFETs with different buildup caps
| Cap | Energy | CR value |
|---|---|---|
| plastic water | 6 MV | 1.000 |
| plastic water | 10 MV | 1.008 |
| brass | 10 MV | 1.038 |
| brass | 6 MV | 0.92 |
It should be noted that for our clinical patient data CR for the 6‐MV beam since the original calibration was done with brass buildup caps as discussed in Section A.
Figure 3MOSFET linearity response for doses ranging from 1 cGy to 300 cGy
Figure 4Dose delivered versus percent standard deviation
Figure 5Correction factors plotted versus SSD
In vivo measurements using MOSFETs compared to corrected target doses from the TPS (CMS XiO). All plans delivered with beam energy 10 MV.
| Patient | Field # |
| IMRT CF | Total CF | Expected MOSFET dose (cGy) | Measured MOSFET dose (cGy) | % Diff |
|---|---|---|---|---|---|---|---|
| 2 | 74.6 | 0.996 | 1.036 | 77.3 | 78.5 | 1.6% | |
| #1 | 3 | 49.0 | 0.992 | 1.049 | 51.4 | 52.6 | 2.2% |
| 4 | 26.7 | 0.975 | 1.020 | 27.3 | 28.7 | 5.1% | |
| 5 | 43.7 | 0.977 | 1.024 | 44.7 | 46.1 | 3.1% | |
| 6 | 84.2 | 1.012 | 1.053 | 88.6 | 91.1 | 2.8% | |
|
|
|
|
|
| |||
| 2 | 61.1 | 1.018 | 1.071 | 65.4 | 65.2 |
| |
| #2 | 3 | 44.7 | 0.996 | 1.009 | 45.1 | 44.0 |
|
| 4 | 34.3 | 1.026 | 1.044 | 35.8 | 35.8 |
| |
| 5 | 39.1 | 1.004 | 1.017 | 39.8 | 39.3 |
| |
| 6 | 62.8 | 0.991 | 1.069 | 67.1 | 70.0 | 4.3% | |
|
|
|
|
|
| |||
| 2 | 74.6 | 0.996 | 1.023 | 76.3 | 77.8 | 1.9% | |
| #3 | 3 | 49.0 | 0.992 | 1.044 | 51.2 | 48.8 |
|
| 4 | 26.7 | 0.975 | 1.027 | 27.5 | 26.3 |
| |
| 5 | 43.7 | 0.977 | 1.044 | 45.6 | 45.7 | 0.3% | |
| 6 | 84.2 | 1.012 | 1.072 | 90.2 | 87.4 |
| |
|
|
|
|
|
| |||
| 2 | 62.9 | 1.001 | 1.039 | 65.3 | 65.1 |
| |
| #4 | 3 | 49.6 | 0.993 | 1.036 | 51.4 | 51.0 |
|
| 4 | 32.0 | 0.953 | 0.998 | 31.9 | 33.2 | 4.0% | |
| 5 | 40.3 | 0.987 | 1.022 | 41.2 | 41.6 | 1.0% | |
| 6 | 62.9 | 1.031 | 1.101 | 69.2 | 65.8 |
| |
|
|
|
|
|
| |||
| 2 | 92.9 | 0.928 | 0.982 | 91.2 | 90.2 |
| |
| #5 | 3 | 51.4 | 0.965 | 0.985 | 50.7 | 52.4 | 3.3% |
| 4 | 31.5 | 0.989 | 1.014 | 31.9 | 30.4 |
| |
| 5 | 50.5 | 1.003 | 1.023 | 51.6 | 51.0 |
| |
| 6 | 92.4 | 0.942 | 1.020 | 94.2 | 103.7 | 10.1% | |
|
|
|
|
|
| |||
| 2 | 73.1 | 0.973 | 1.010 | 73.8 | 75.0 | 1.6% | |
| #6 | 3 | 52.7 | 0.976 | 0.996 | 52.6 | 51.4 |
|
| 4 | 22.5 | 0.928 | 0.954 | 21.4 | 22.6 | 5.6% | |
| 5 | 45.7 | 0.964 | 0.984 | 45.0 | 47.0 | 4.4% | |
| 6 | 78.3 | 0.980 | 1.036 | 81.1 | 85.1 | 5.0% | |
|
|
|
|
|
|
In vivo measurements using MOSFETs compared to corrected target doses from the TPS (Philips Pinnacle ). All plans delivered with beam energy 6 MV.
| Patient | Field # |
| IMRT CF | Total CF | Expected MOSFET dose (cGy) | Measured MOSFET dose (cGy) | % Diff |
|---|---|---|---|---|---|---|---|
| 2 | 82.5 | 0.971 | 0.984 | 81.2 | 87.4 | 7.6% | |
| #7 | 3 | 53.2 | 1.006 | 1.027 | 54.6 | 54.0 | 1.2% |
| 4 | 41.8 | 0.995 | 1.018 | 42.6 | 42.6 | 0.1% | |
| 5 | 54.2 | 0.984 | 1.005 | 54.5 | 56.3 | 3.4% | |
| 6 | 84.0 | 0.976 | 0.988 | 83.1 | 87.4 | 5.2% | |
|
|
|
|
|
| |||
| 2 | 105.0 | 0.983 | 0.989 | 103.9 | 109.0 | 4.9% | |
| #8 | 3 | 49.9 | 0.973 | 1.035 | 51.6 | 53.2 | 3.0% |
| 4 | 36.0 | 0.979 | 1.022 | 36.8 | 35.4 | 3.8% | |
| 5 | 48.5 | 0.971 | 1.032 | 50.1 | 52.6 | 5.1% | |
| 6 | 91.6 | 0.972 | 0.985 | 90.2 | 93.5 | 3.6% | |
|
|
|
|
|
| |||
| 2 | 112.0 | 0.976 | 0.976 | 109.4 | 115.0 | 5.2% | |
| #9 | 3 | 46.3 | 0.951 | 1.006 | 46.6 | 48.2 | 3.5% |
| 4 | 21.9 | 0.944 | 0.983 | 21.5 | 22.7 | 5.4% | |
| 5 | 59.3 | 0.965 | 1.009 | 59.8 | 63.0 | 5.3% | |
| 6 | 118.3 | 0.967 | 1.073 | 126.9 | 121.0 | 4.7% | |
|
|
|
|
|
| |||
| #10 | 2 | 40.2 | 0.981 | 1.022 | 41.1 | 39.3 | 4.4% |
| prosthesis | 3 | 54.0 | 0.932 | 0.960 | 51.8 | 52.2 | 0.7% |
| (#6 not | 4 | 66.8 | 0.966 | 0.977 | 65.3 | 65.7 | 0.6% |
| measured) | 5 | 89.2 | 0.951 | 0.975 | 87.0 | 88.0 | 1.2% |
|
|
|
|
|
| |||
| 2 | 81.5 | 0.958 | 0.961 | 78.3 | 80.4 | 2.6% | |
| #11 | 3 | 72.8 | 0.971 | 0.991 | 72.1 | 68.3 | 5.3% |
| 4 | 43.5 | 0.957 | 0.979 | 42.6 | 41.5 | 2.6% | |
| 5 | 70.1 | 0.984 | 1.032 | 72.4 | 68.8 | 4.9% | |
| 6 | 82.4 | 0.982 | 1.009 | 83.2 | 86.4 | 3.9% | |
|
|
|
|
|
|
Uncertainty of the MOSFET dose for IMRT therapy given as 1 SD (high sensitivity)
| Uncertainty (%) | |
|---|---|
|
| |
| Dose measured with ionization chamber | 1.0% |
| Reading of the MOSFET detector | 1.3% |
| Reproducibility of the setting | 0.5% |
| Combined uncertainty of step 1 | 1.7% |
|
| |
| SSD correction factor | 1.4% |
| CR value | 1.5% |
| IMRT correction factor–MOSFET reading on solid water phantom | 1.3% |
| Combined uncertainty of step 2 | 2.4% |
|
| |
| Reading of the MOSFET detector | 1.3% |
| Positioning (2% over 2 mm radius) | 1.5% |
| MOSFET detector long‐term stability over time | 3.0% |
| Combined uncertainty of step 3 | 3.6% |
|
|
|