Interventional MRI at high-field (1.5 T): needle artifacts.

H Liu1, A J Martin, C L Truwit.   

Abstract

A better understanding of the appearance of a biopsy needle as well as its interaction with various sequence parameters in MRI is beneficial for its application in interventional MRI. As an extension of previous researchers' contributions, we investigate the specific characteristics of MR image artifacts associated with the tip of a biopsy needle when it is approximately parallel to the main magnetic field. The origin of the needle tip artifact, which exhibits as a blooming ball, was studied using MRI techniques and numerical simulation using the finite element method (FEM). Satisfactory agreement between theory and experiment has been achieved. Results showed that the image artifacts associated with biopsy needle are present and dependent on imaging parameters, but the artifacts can be reduced if optimal imaging parameters were used. Images of actual human brain tumor biopsies performed using the same needle under MRI guidance and monitoring demonstrated this artifact.

Entities:  

Mesh:

Year:  1998        PMID: 9500283     DOI: 10.1002/jmri.1880080136

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  7 in total

1.  System for MR image-guided prostate interventions: canine study.

Authors:  Robert C Susil; Axel Krieger; J Andrew Derbyshire; Attila Tanacs; Louis L Whitcomb; Gabor Fichtinger; Ergin Atalar
Journal:  Radiology       Date:  2003-09       Impact factor: 11.105

2.  Magnetic resonance-guided upper abdominal biopsies in a high-field wide-bore 3-T MRI system: feasibility, handling, and needle artefacts.

Authors:  Jens-Peter Kühn; Sönke Langner; Katrin Hegenscheid; Matthias Evert; Antje Kickhefel; Norbert Hosten; Ralf Puls
Journal:  Eur Radiol       Date:  2010-05-26       Impact factor: 5.315

3.  Numerical simulations of intra-voxel dephasing effects and signal voids in gradient echo MR imaging using different sub-grid sizes.

Authors:  Bernd M Müller-Bierl; Hansjörg Graf; Philippe L Pereira; Fritz Schick
Journal:  MAGMA       Date:  2006-05-09       Impact factor: 2.310

4.  Deblurring for spiral real-time MRI using convolutional neural networks.

Authors:  Yongwan Lim; Yannick Bliesener; Shrikanth Narayanan; Krishna S Nayak
Journal:  Magn Reson Med       Date:  2020-07-25       Impact factor: 4.668

Review 5.  Limitations of Resting-State Functional MR Imaging in the Setting of Focal Brain Lesions.

Authors:  Shruti Agarwal; Haris I Sair; Jay J Pillai
Journal:  Neuroimaging Clin N Am       Date:  2017-11       Impact factor: 2.264

6.  Fast Fourier-based simulation of off-resonance artifacts in steady-state gradient echo MRI applied to metal object localization.

Authors:  Frank Zijlstra; Job G Bouwman; Ieva Braškutė; Max A Viergever; Peter R Seevinck
Journal:  Magn Reson Med       Date:  2016-12-07       Impact factor: 4.668

7.  Magnetic resonance imaging-guided biopsies in children.

Authors:  Mika Hirvonen; Juha-Jaakko Sinikumpu; Osmo Tervonen; Roberto Blanco Sequeiros
Journal:  Acta Radiol Open       Date:  2021-11-29
  7 in total