Abstract

OBJECTIVE: Evaluate the accuracy of this bedside method to determine hemoglobin (Hb) concentration in general surgery over a wide range of Hb values and to determine potential sources of error.
METHODS: Accuracy of Hb measurement using HemoCue (AB Leo Diagnostics, Helsinborg, Sweden) was assessed in 140 surgical blood samples using 7 HemoCue devices in comparison with a CO-Oximeter (IL 482, Instrumentation Laboratory, Lexington, MA). To analyze potential sources of error, packed red cells and fresh frozen plasma were reconstituted to randomized Hb levels of 2-18 g/dL.
RESULTS: In the surgical blood samples, the Hb concentration determined by the CO-Oximeter (HbCOOX) ranged from 5.1 to 16.7 g/dL and the Hb concentration measured by HemoCue (HbHC) from 4.7 to 16.0 g/dL. Bias (HbCOOX - HbHC) between HbCOOX and HbHC was 0.6+/-0.6 g/dL (mean +/- SD) or 5.4+/-5.0% (p < 0.001). Also in the reconstituted blood, the bias between HbCOOX and HbHC was significant (0.2+/-0.3 g/dL or 2.1+/-3.2%; p < 0.001). The microcuvette explained 68% of the variability between HbCOOX and HbHC. HemoCue thus underestimates the Hb concentration by 2-5% and exhibits a 8-10 times higher variability with only 86.4% of HbHC being within +/- 10% of HbCOOX. CONCLUSION. Although the mean bias between HbCOOX and HbHC was relatively low, Hb measurement by HemoCue exhibited a significant variability. Loading multiple microcuvettes and averaging the results may increase the accuracy of Hb measurement by HemoCue.

Mesh:

Substances:

Year:  1997        PMID: 9495289     DOI: 10.1023/a:1007451611748

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  14 in total

1.  A clinical evaluation of the Hemocue haemoglobinometer using capillary, venous and arterial samples.

Authors:  P P Chen; T G Short; D H Leung; T E Oh
Journal:  Anaesth Intensive Care       Date:  1992-11       Impact factor: 1.669

2.  Rapid assessment of fetal hemoglobin concentration with the HemoCue system.

Authors:  S A Laifer; J A Kuller; L M Hill
Journal:  Obstet Gynecol       Date:  1990-10       Impact factor: 7.661

3.  Hemocue--a new haemoglobinometer in the clinic.

Authors:  W K Be; H E Kerkkamp; L H Booij
Journal:  Eur J Anaesthesiol       Date:  1991-01       Impact factor: 4.330

4.  Fetal hemoglobin quantitations using the Hemocue system are rapid and accurate.

Authors:  S M Berry; M P Dombrowski; W B Blessed; J A Bichalski; T B Jones; D B Cotton
Journal:  Obstet Gynecol       Date:  1993-03       Impact factor: 7.661

5.  Evaluation of portable haemoglobinometer in general practice.

Authors:  R G Neville
Journal:  Br Med J (Clin Res Ed)       Date:  1987-05-16

6.  Evaluation of the IL 482 CO-Oximeter.

Authors:  F A Sedor; C M Holleman
Journal:  Clin Chem       Date:  1987-10       Impact factor: 8.327

7.  High humidity affects HemoCue microcuvette function.

Authors:  M A Henderson; M G Irwin
Journal:  Anaesth Intensive Care       Date:  1995-06       Impact factor: 1.669

8.  Autologous blood donation--can we bank on it?

Authors:  C J Rutherford; H S Kaplan
Journal:  N Engl J Med       Date:  1995-03-16       Impact factor: 91.245

9.  HemoCue system for hemoglobin measurement. Evaluation in anemic and nonanemic children.

Authors:  A R Cohen; J Seidl-Friedman
Journal:  Am J Clin Pathol       Date:  1988-09       Impact factor: 2.493

10.  A comparison of four bedside methods of hemoglobin assessment during cardiac surgery.

Authors:  S E McNulty; M Torjman; W Grodecki; A Marr; H Schieren
Journal:  Anesth Analg       Date:  1995-12       Impact factor: 5.108

View more
  14 in total

1.  Comparison of HemoCue® hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan.

Authors:  Ishag Adam; Samah Ahmed; Mahmoud H Mahmoud; Mohammed I Yassin
Journal:  Diagn Pathol       Date:  2012-03-21       Impact factor: 2.644

2.  Reliability of HemoCue in patients with gastrointestinal bleeding.

Authors:  Andry Van de Louw; Nadine Lasserre; François Drouhin; Stéphane Thierry; Lucien Lecuyer; Daniel Caen; Alain Tenaillon
Journal:  Intensive Care Med       Date:  2006-12-05       Impact factor: 17.440

Review 3.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

Review 4.  Prevalence of Anaemia in Kerala State, Southern India - A Systematic Review.

Authors:  PuruShothama Suseela Rakesh
Journal:  J Clin Diagn Res       Date:  2017-05-01

5.  Evaluation of a portable hemoglobin photometer in pregnant women in a high altitude area: a pilot study.

Authors:  Xiaoyan Zhou; Hong Yan; Yuan Xing; Shaonong Dang; Bianba Zhuoma; Duolao Wang
Journal:  BMC Public Health       Date:  2009-07-11       Impact factor: 3.295

6.  Hepcidin/Ferritin Quotient Helps to Predict Spontaneous Recovery from Iron Loss following Blood Donation.

Authors:  Ramin Lotfi; Christine Kroll; Dietmar Plonné; Bernd Jahrsdörfer; Hubert Schrezenmeier
Journal:  Transfus Med Hemother       Date:  2015-11-03       Impact factor: 3.747

7.  Hemoglobin estimation by the HemoCue® portable hemoglobin photometer in a resource poor setting.

Authors:  Bernard Nkrumah; Samuel Blay Nguah; Nimako Sarpong; Denise Dekker; Ali Idriss; Juergen May; Yaw Adu-Sarkodie
Journal:  BMC Clin Pathol       Date:  2011-04-21

8.  Acute iatrogenic polycythemia induced by massive red blood cell transfusion during subtotal abdominal colectomy.

Authors:  David Chiapaikeo; Pejman Rohani
Journal:  Hematol Rep       Date:  2015-03-23

9.  Influence of age on the haemoglobin concentration of malaria-infected patients in a reference centre in the Brazilian Amazon.

Authors:  Andre M Siqueira; Janieldo A Cavalcante; Shelia Vítor-Silva; Roberto C Reyes-Lecca; Aline C Alencar; Wuelton M Monteiro; Márcia A A Alexandre; Maria Paula G Mourão; Caterina Guinovart; Quique Bassat; Maria das Graças C Alecrim; Marcus V G Lacerda
Journal:  Mem Inst Oswaldo Cruz       Date:  2014-08-19       Impact factor: 2.743

10.  Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study.

Authors:  Wael N Awada; Maher F Mohmoued; Tarek M Radwan; Gomaa Z Hussien; Hany W Elkady
Journal:  J Clin Monit Comput       Date:  2015-02-04       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.