| Literature DB >> 9399498 |
S Ichiyama1, Y Ito, F Sugiura, Y Iinuma, S Yamori, M Shimojima, Y Hasegawa, K Shimokata, N Nakashima.
Abstract
We compared the ability of the semiautomated BDProbeTec-SDA system, which uses the strand displacement amplification (SDA) method, with that of the Roche Amplicor-PCR system and the Septi-Chek AFB culture system to directly detect Mycobacterium tuberculosis complex (MTB) and other mycobacteria in sputum samples. A total of 530 sputum samples from 299 patients were examined in this study. Of the 530 samples, 129 were culture positive for acid-fast bacilli with the Septi-Chek AFB system; 95 for MTB, 29 for M. avium-M. intracellulare complex (MAC), and 5 for other mycobacteria. The BDProbeTec-SDA system detected 90 of the 95 samples culture positive for MTB (sensitivity, 94.7%), and the Amplicor-PCR system detected 85 of the 95 samples culture positive for MTB (sensitivity, 89.5%). The specificity of each system, based on the clinical diagnosis, was 99.8% for SDA and 100% for PCR, respectively. Among the 29 samples culture positive for MAC, the BDProbeTec-SDA system detected MAC in 24 samples (sensitivity, 82.8%), whereas the Amplicor-PCR system detected MAC in 23 samples (sensitivity, 79.3%). The specificities of the systems were 98.3 and 100%, respectively. The high degrees of sensitivity and specificity of the BDProbeTec-SDA system suggest that it should be very useful in clinical laboratories for the rapid detection of mycobacteria in sputum samples.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9399498 PMCID: PMC230126 DOI: 10.1128/jcm.35.12.3082-3085.1997
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948