Adverse anesthetic outcomes arising from gas delivery equipment: a closed claims analysis.

R A Caplan1, M F Vistica, K L Posner, F W Cheney.   

Abstract

BACKGROUND: Anesthesia gas delivery equipment is a potentially important source of patient injury. To better define the contribution of gas delivery equipment to professional liability in anesthesia, the authors conducted an in-depth analysis of cases from the database of the American Society of Anesthesiologists Closed Claims Project.
METHODS: The database of the Closed Claims Project is composed of closed US malpractice claims that have been collected in a standardized manner. All claims resulting from the use of gas delivery equipment were reviewed for recurrent patterns of injury.
RESULTS: Gas delivery equipment was associated with 72 (2%) of 3,791 claims in the database. Death and permanent brain damage accounted for almost all adverse outcomes (n = 55, 76%). Equipment misuse was defined as fault or human error associated with the preparation, maintenance, or deployment of a medical device. Equipment failure was defined as unexpected malfunction of a medical device, despite routine maintenance and previous uneventful use. Misuse of equipment (n = 54, 75%) was three times more common than equipment failure (n = 17, 24%). Misconnects and disconnects of the breathing circuit made the largest contribution to injury (n = 25, 35%). Reviewers judged that 38 of 72 claims (53%) could have been prevented by pulse oximetry, capnography, or a combination of these two monitors. Overall, 56 of 72 gas delivery claims (78%) were deemed preventable with the use or better use of monitors. The year of occurrence for claims involving gas delivery equipment ranged from 1962 to 1991 and did not differ significantly from claims involving other adverse respiratory events.
CONCLUSIONS: Claims associated with gas delivery equipment are infrequent but severe and continue to occur in the 1990s. Educational and preventive strategies that focus on equipment misuse and breathing circuit configuration may have the greatest potential for enhancing the safety of anesthesia gas delivery equipment.

Entities:  

Mesh:

Year:  1997        PMID: 9357874     DOI: 10.1097/00000542-199710000-00006

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  17 in total

1.  Preoperative anaesthesia information--what do patients need to know?

Authors:  P Lennox; A J Cunningham
Journal:  Ir J Med Sci       Date:  2000 Apr-Jun       Impact factor: 1.568

2.  Medical school attended as a predictor of medical malpractice claims.

Authors:  T M Waters; F V Lefevre; P P Budetti
Journal:  Qual Saf Health Care       Date:  2003-10

3.  Monitoring of breathing phases using a bioacoustic method in healthy awake subjects.

Authors:  Hisham Alshaer; Geoffrey R Fernie; T Douglas Bradley
Journal:  J Clin Monit Comput       Date:  2011-09-29       Impact factor: 2.502

4.  The Importance of State and Context in Safe Interoperable Medical Systems.

Authors:  Sandy Weininger; Michael B Jaffe; Michael Robkin; Tracy Rausch; David Arney; Julian M Goldman
Journal:  IEEE J Transl Eng Health Med       Date:  2016-08-08       Impact factor: 3.316

5.  FLOW-i ventilator performance in the presence of a circle system leak.

Authors:  Umberto Lucangelo; Miloš Ajčević; Agostino Accardo; Massimo Borelli; Alberto Peratoner; Lucia Comuzzi; Walter A Zin
Journal:  J Clin Monit Comput       Date:  2016-04-09       Impact factor: 2.502

6.  Accidental oxygen disconnection in the emergency department.

Authors:  Guyon J Hill; Bruce D Adams
Journal:  J Emerg Trauma Shock       Date:  2010-04

Review 7.  [Perioperative risk and mortality after major surgery].

Authors:  O Boehm; M K A Pfeiffer; G Baumgarten; A Hoeft
Journal:  Anaesthesist       Date:  2015-11       Impact factor: 1.041

Review 8.  Risks from Catheter Ablation of Atrial Fibrillation: A Review of Methods, Efficacy, and Safety.

Authors:  Richard Cheng; Melisa Chang; Chee Yuan Ng
Journal:  J Atr Fibrillation       Date:  2012-04-14

9.  The correlation and level of agreement between end-tidal and blood gas pCO2 in children with respiratory distress: a retrospective analysis.

Authors:  James M Moses; Jamin L Alexander; Michael Sd Agus
Journal:  BMC Pediatr       Date:  2009-03-12       Impact factor: 2.125

Review 10.  The modern integrated anaesthesia workstation.

Authors:  Vijaya P Patil; Madhavi G Shetmahajan; Jigeeshu V Divatia
Journal:  Indian J Anaesth       Date:  2013-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.