Combined use of non-thoracotomy cardioverter defibrillators and endocardial pacemakers.

H H Noguera1, A O Peralta, R M John, F J Venditti, D T Martin.   

Abstract

OBJECTIVE: To study the potential interactions in patients with endocardial permanent pacemakers and non-thoracotomy implantable cardioverter defibrillator (ICD) systems.
DESIGN: Case series and cohort study.
SETTING: Tertiary referral centre. PATIENTS: Fifteen consecutive patients with both endocardial pacemakers (12 dual chamber and three single chamber) and non-thoracotomy ICD systems. MAIN OUTCOME MEASURES: Detection inhibition of induced ventricular fibrillation; double counting; and pacemaker function after shocks. In the evaluation of detection inhibition, 124 VF inductions were analysed for detection duration compared with induced VF episodes in controls with an ICD but without a pacemaker.
RESULTS: Two patients (13%) showed detection inhibition of VF and required pacemaker system change at the time of the ICD implant. With the final lead position, despite frequent pacemaker undersensing of VF, ICD detection of VF was not inhibited during any induction, and neither initial detection nor redetection times for VF were different from controls. Double/triple counting of pacemaker artefact and evoked electrogram was noted in three patients (20%). In two, this was remedied during the implantation procedure, and in the other it was abolished when amiodarone treatment was discontinued. Pacemaker function was affected by ICD discharges in two patients, one who showed postshock atrial undersensing and loss of capture, and another whose pacemaker reverted to VVI mode.
CONCLUSIONS: When careful testing is performed at implantation to detect and remedy device interactions, non-thoracotomy ICD treatment and endocardial pacemakers can be used safely in combination.

Entities:  

Mesh:

Year:  1997        PMID: 9290402      PMCID: PMC484864          DOI: 10.1136/hrt.78.1.50

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  11 in total

1.  The right ventricular outflow tract as an alternative permanent pacing site: long-term follow-up.

Authors:  E S Barin; S M Jones; D E Ward; A J Camm; A W Nathan
Journal:  Pacing Clin Electrophysiol       Date:  1991-01       Impact factor: 1.976

2.  Combined automatic implantable cardioverter-defibrillator and pacemaker systems: implantation techniques and follow-up.

Authors:  A E Epstein; G N Kay; V J Plumb; R B Shepard; J K Kirklin
Journal:  J Am Coll Cardiol       Date:  1989-01       Impact factor: 24.094

3.  The use and interaction of permanent pacemakers and the automatic implantable cardioverter defibrillator.

Authors:  A I Cohen; M H Wish; R D Fletcher; F C Miller; D McCormick; J Shuck; N Shapira; A A Delnegro
Journal:  Pacing Clin Electrophysiol       Date:  1988-06       Impact factor: 1.976

4.  Unipolar pacer artifacts induced failure of an automatic implantable cardioverter/defibrillator to detect ventricular fibrillation.

Authors:  S G Kim; S Furman; L E Waspe; R Brodman; J D Fisher
Journal:  Am J Cardiol       Date:  1986-04-01       Impact factor: 2.778

5.  Combined automatic implantable cardioverter-defibrillator and pacemaker systems...

Authors:  S Furman
Journal:  J Am Coll Cardiol       Date:  1989-01       Impact factor: 24.094

6.  Safety of pacemaker implantation in patients with transvenous (nonthoracotomy) implantable cardioverter defibrillators.

Authors:  H F Clemo; K A Ellenbogen; M K Belz; M A Wood; B S Stambler
Journal:  Pacing Clin Electrophysiol       Date:  1994-12       Impact factor: 1.976

7.  Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators.

Authors:  H Calkins; J Brinker; E P Veltri; T Guarnieri; J H Levine
Journal:  J Am Coll Cardiol       Date:  1990-09       Impact factor: 24.094

8.  Feasibility of concomitant implantation of permanent transvenous pacemaker and defibrillator systems.

Authors:  Z Blanck; I Niazi; K Axtell; J Sra; M R Jazayeri; A Dhala; S Deshpande; M Akhtar
Journal:  Am J Cardiol       Date:  1994-12-15       Impact factor: 2.778

9.  Cephalic vein guide wire technique for implantation of permanent pacemakers.

Authors:  L S Ong; S S Barold; M Lederman; M D Falkoff; R A Heinle
Journal:  Am Heart J       Date:  1987-10       Impact factor: 4.749

10.  Implantation of transvenous nonthoracotomy cardioverter-defibrillator systems in patients with permanent endocardial pacemakers.

Authors:  R Brooks; H Garan; B A McGovern; J N Ruskin
Journal:  Am Heart J       Date:  1995-01       Impact factor: 4.749

View more
  1 in total

1.  Inappropriate implantable defibrillator discharge caused by a retained pacemaker lead fragment.

Authors:  L Lickfett; C Wolpert; W Jung; S Spehl; L Pizzulli; B Esmailzadeh; B Lüderitz
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.