The use and interaction of permanent pacemakers and the automatic implantable cardioverter defibrillator.

A I Cohen1, M H Wish, R D Fletcher, F C Miller, D McCormick, J Shuck, N Shapira, A A Delnegro.   

Abstract

The adverse interactions of permanent pacemakers and automatic implantable cardioverter defibrillators (AICD) were studied in nine patients in whom both devices were implanted. Both unipolar and bipolar pacemakers were evaluated. The permanent pacemakers were also used to do noninvasive electrophysiological studies and to induce ventricular fibrillation. Undersensing of ventricular fibrillation by the permanent pacemakers caused inappropriate pacemaker stimuli, which caused undersensing of ventricular fibrillation by the AICD in three of four patients with unipolar pacemakers. After an AICD discharge, pacemaker noncapture was seen in eight of 22 episodes for an average 4.9 seconds and inability to sense was seen in 11 of 20 episodes for an average 9.0 seconds. Counting of pacemaker stimuli and QRS by the AICD caused inappropriate discharges. Noninvasive electrophysiological testing by the pacemakers correlated with invasive testing. Furthermore, induction of ventricular fibrillation was successful in four of five patients attempted, though requiring long bursts at high outputs at the shortest cycle lengths obtainable by these pacemakers. Operation of the AICD and permanent pacemakers must be clearly understood to avoid adverse interactions of these devices.

Entities:  

Mesh:

Year:  1988        PMID: 2456550     DOI: 10.1111/j.1540-8159.1988.tb06020.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Combined use of non-thoracotomy cardioverter defibrillators and endocardial pacemakers.

Authors:  H H Noguera; A O Peralta; R M John; F J Venditti; D T Martin
Journal:  Heart       Date:  1997-07       Impact factor: 5.994

2.  Use of an implantable cardioverter defibrillator in a patient with a rate responsive pacemaker.

Authors:  R Chamberlain-Webber; I Rankin; R Sutton
Journal:  Br Heart J       Date:  1994-02

3.  Unique interaction between an atrial single-chamber pacemaker and a ventricular defibrillator.

Authors:  C Sticherling; S A Strickberger; G F Michaud
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

4.  [Clinical experience with combined automatic implantable cardioverter-defibrillator and pacemaker systems].

Authors:  B Hailer; U Wolfhard; K J Altmaier; M Wehr
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.