The effect of cuff width on systolic blood pressure measurements in neonates.

L G Lum, M D Jones.   

Abstract

The width of the blood pressure cuff was found to have an effect on the accuracy of systolic blood pressure measured in the arm and thigh in neonates. Data were collected in 26 infants of a variety of birth weights and gestational ages. They show that the effect of cuff size can be minimized by selecting a cuff width equal to approximately 50% of the extremity circumference.

Entities:  

Mesh:

Year:  1977        PMID: 925831     DOI: 10.1016/s0022-3476(77)80905-0

Source DB:  PubMed          Journal:  J Pediatr        ISSN: 0022-3476            Impact factor:   4.406


  5 in total

Review 1.  Renal disorders and diseases of the newborn.

Authors:  E John
Journal:  Indian J Pediatr       Date:  1985 Jan-Feb       Impact factor: 1.967

2.  Non-invasive oscillometric measurement of systolic, mean and diastolic blood pressure in infants with congenital heart defects after operation. A comparison with direct blood pressure measurements.

Authors:  V Pilossoff; J G Schöber; D Peters; K Bühlmeyer
Journal:  Eur J Pediatr       Date:  1985-11       Impact factor: 3.183

3.  Rapid measurement of blood pressure in the sick neonate.

Authors:  E S Ogata; G R Herring
Journal:  Intensive Care Med       Date:  1980       Impact factor: 17.440

4.  Systolic blood pressure in babies of less than 32 weeks gestation in the first year of life. Northern Neonatal Nursing Initiative.

Authors: 
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-01       Impact factor: 5.747

5.  Mean mid-arm circumference and blood pressure cuff sizes for US children, adolescents and adults: National Health and Nutrition Examination Survey, 2011-2016.

Authors:  Yechiam Ostchega; Jeffery P Hughes; Tatiana Nwankwo; Guangyu Zhang
Journal:  Blood Press Monit       Date:  2018-12       Impact factor: 1.444

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.