Electrocardiogram-gated intravascular ultrasound image acquisition after coronary stent deployment facilitates on-line three-dimensional reconstruction and automated lumen quantification.

C von Birgelen1, G S Mintz, A Nicosia, D P Foley, W J van der Giessen, N Bruining, S G Airiian, J R Roelandt, P J de Feyter, P W Serruys.   

Abstract

OBJECTIVE: This study evaluates the feasibility, reliability and reproducibility of electrocardiogram (ECG)-gated intravascular ultrasound (IVUS) image acquisition during automated transducer withdrawal and automated three-dimensional (3D) boundary detection for assessing on-line the result of coronary stenting.
BACKGROUND: Systolic-diastolic image artifacts frequently limit the clinical applicability of such automated analysis systems.
METHODS: In 30 patients, after successful angiography-guided implantation of 34 stents in 30 target lesions, we carried out IVUS examinations on-line with the use of ECG-gated automated 3D analyses and conventional manual analyses of two-dimensional images from continuous pullbacks. These on-line measurements were compared with off-line 3D reanalyses. The adequacy of stent deployment was determined by using ultrasound criteria for stent apposition, symmetry and expansion.
RESULTS: Gated image acquisition was successfully performed in all patients to allow on-line 3D analysis within 8.7 +/- 0.6 min (mean +/- SD). Measurements by on-line and off-line 3D analyses correlated closely (r > or = 0.95), and the minimal stent lumen differed only minimally (8.6 +/- 2.8 mm2 vs. 8.5 +/- 2.8 mm2, p = NS). The conventional analysis significantly overestimated the minimal stent lumen (9.0 +/- 2.7 mm2, p < 0.005) in comparison with results of both 3D analyses. Fourteen stents (41%) failed to meet the criteria by both 3D analyses, all of these not reaching optimal expansion, but only 7 (21%) were detected by conventional analysis (p < 0.02). Intraobserver and interobserver comparison of stent lumen measurements by the automated approach revealed minimal differences (0.0 +/- 0.2 mm2 and 0.0 +/- 0.3 mm2) and excellent correlations (r = 0.99 and 0.98, respectively).
CONCLUSIONS: ECG-gated image acquisition after coronary stent deployment is feasible, permits on-line automated 3D reconstruction and analysis and provides reliable and reproducible measurements; these factors facilitate detection of the minimal lumen site.

Entities:  

Mesh:

Year:  1997        PMID: 9247516     DOI: 10.1016/s0735-1097(97)00154-x

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  9 in total

Review 1.  Invasive assessment of the coronary circulation: intravascular ultrasound and Doppler.

Authors:  David E Newby; Keith A A Fox
Journal:  Br J Clin Pharmacol       Date:  2002-06       Impact factor: 4.335

2.  Validation of an automated system for luminal and medial-adventitial border detection in three-dimensional intravascular ultrasound.

Authors:  Jon D Klingensmith; E Murat Tuzcu; Steven E Nissen; D Geoffrey Vince
Journal:  Int J Cardiovasc Imaging       Date:  2003-04       Impact factor: 2.357

3.  Quantitative assessment of mild-to-moderate coronary atherosclerosis by computerized border detection in 3D IVUS.

Authors:  Clemen von Birgelen
Journal:  Int J Cardiovasc Imaging       Date:  2003-04       Impact factor: 2.357

4.  Evaluation of the influence of cardiac motion on the accuracy and reproducibility of longitudinal measurements and the corresponding image quality in optical frequency domain imaging: an ex vivo investigation of the optimal pullback speed.

Authors:  Kohei Koyama; Kihei Yoneyama; Takanobu Mitarai; Shingo Kuwata; Ken Kongoji; Tomoo Harada; Yoshihiro J Akashi
Journal:  Int J Cardiovasc Imaging       Date:  2015-05-14       Impact factor: 2.357

5.  Single cardiac cycle three-dimensional intracoronary optical coherence tomography.

Authors:  Tae Shik Kim; Hyun-Sang Park; Sun-Joo Jang; Joon Woo Song; Han Saem Cho; Sunwon Kim; Brett E Bouma; Jin Won Kim; Wang-Yuhl Oh
Journal:  Biomed Opt Express       Date:  2016-11-01       Impact factor: 3.732

6.  Coronary stent symmetry and vascular injury determine experimental restenosis.

Authors:  C Schulz; R A Herrmann; C Beilharz; J Pasquantonio; E Alt
Journal:  Heart       Date:  2000-04       Impact factor: 5.994

7.  Atherosclerotic coronary lesions with inadequate compensatory enlargement have smaller plaque and vessel volumes: observations with three dimensional intravascular ultrasound in vivo.

Authors:  C von Birgelen; G S Mintz; E A de Vrey; T Kimura; J J Popma; S G Airiian; M B Leon; M Nobuyoshi; P W Serruys; P J de Feyter
Journal:  Heart       Date:  1998-02       Impact factor: 5.994

8.  Intravascular molecular-structural imaging with a miniaturized integrated near-infrared fluorescence and ultrasound catheter.

Authors:  Stephan Kellnberger; Georg Wissmeyer; Mazen Albaghdadi; Zhonglie Piao; Wenzhu Li; Adam Mauskapf; Philipp Rauschendorfer; Guillermo J Tearney; Vasilis Ntziachristos; Farouc A Jaffer
Journal:  J Biophotonics       Date:  2021-07-20       Impact factor: 3.390

Review 9.  The Evolution of Data Fusion Methodologies Developed to Reconstruct Coronary Artery Geometry From Intravascular Imaging and Coronary Angiography Data: A Comprehensive Review.

Authors:  Yakup Kilic; Hannah Safi; Retesh Bajaj; Patrick W Serruys; Pieter Kitslaar; Anantharaman Ramasamy; Vincenzo Tufaro; Yoshinobu Onuma; Anthony Mathur; Ryo Torii; Andreas Baumbach; Christos V Bourantas
Journal:  Front Cardiovasc Med       Date:  2020-03-31
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.