Abstract
When an electrical stimulus is used to evoke action potentials in peripheral nerves or the spinal cord, the stimulus causes an artefact which may interfere with measurement of the evoked potentials. This artefact, unlike all other sources of noise in the measurement, cannot be reduced by ensemble averaging. Confusion about the origin and transmission of stimulus artefact has led to considerable frustration in spinal somatosensory evoked potential (SSEP) measurements. The three mechanisms by which stimulus artefact is coupled into the measuring system are identified, and means for their reduction are discussed.Mesh:
Year: 1997 PMID: 9246854 DOI: 10.1007/bf02530040
Source DB: PubMed Journal: Med Biol Eng Comput ISSN: 0140-0118 Impact factor: 2.602