A Jivan1, M A Horsfield, A R Moody, G R Cherryman.
Abstract
The application of an inversion-recovery snapshot FLASH (fast low-angled shot) imaging sequence to the dynamic measurement of monoexponential T1 relaxation was investigated. The effect of (a) a reduction in the overall sequence repetition time, and (b) an increase of the read-pulse flip angle, on the measurement of T1 was analyzed. The error in T1 introduced by these factors is calculated, and a fuller analysis that takes them into account is presented. Data from a phantom are used to confirm this analysis. The magnitude of the errors is illustrated by measuring myocardial T1 in patients with acute ischaemic heart disease during the injection of a bolus of the contrast medium gadobenate dimeglumine. Overall, there was a 10% difference between the T1 values when the approximate and exact solutions were used; this was statistically significant. However, the difference was on average 25% for patients with a high heart rate (because of the shorter sequence-repetition time) in areas of infarcted myocardium (because of the longer T1).Entities:
Mesh:
Substances:
Year: 1997 PMID: 9245631 DOI: 10.1006/jmre.1997.1177
Source DB: PubMed Journal: J Magn Reson ISSN: 1090-7807 Impact factor: 2.229