The intrastromal corneal ring segments. Phase II results for the correction of myopia.

D J Schanzlin1, P A Asbell, T E Burris, D S Durrie.   

Abstract

OBJECTIVE: The purpose of the study was to evaluate the safety and efficacy of the intrastromal corneal ring segments (ICRS) for the correction of myopia.
DESIGN: A 2-year phase II clinical trial of ICRS was initiated in May 1995. The investigational plan specifies that 150 patients with sighted eyes, requiring myopic corrections from -1.00 to -6.00 diopters (D), will each receive ICRS in 1 eye. The patient population will be divided into approximately five patients per ICRS thickness (0.25, 0.30, 0.35, 0.40, and 0.45 mm) per site. Six investigational sites are participating in the trial. PARTICIPANTS: Fifty-nine men and 43 women requiring myopic corrections were enrolled at four U.S. investigational sites. These 102 patients each received the ICRS product in 1 eye. INTERVENTION: Correction of myopia. MAIN OUTCOME MEASURES: Efficacy of ICRS was assessed with respect to the trial endpoints of predictability of refractive effect, uncorrected visual acuity (UCVA), stability of UCVA, maintenance of best spectacle-corrected visual acuity and stability of refractive effect.
RESULTS: As shown by the available month-3 data (99 patients; all device thicknesses), 95 (96%) of 99 patients had a UCVA of 20/40 or better. Ninety-eight (99%) of 99 patients were within 2 lines of their preoperative best spectacle-corrected visual acuity. The average change (with standard error) in cycloplegic refraction (spherical equivalent) achieved by ICRS thickness was -1.27 +/- 0.09 D (0.25 mm), -2.13 +/- 0.16 D (0.30 mm), -2.56 +/- 0.15 D (0.35 mm), -3.77 +/- 0.37 D (0.40 mm) and -4.16 +/- 0.24 D (0.45 mm). Seventy-seven percent (76/99) of the patients were within +/-1.00 D of their intended correction. When the ICRS was removed in two cases, both patients returned to within 0.75 D of their preoperative manifest refraction.
CONCLUSIONS: The ICRS appears to be a viable and effective alternative for the treatment of myopia. Additionally, as indicated by the explant data, the ICRS's refractive effect may be reversible upon removal of the device.

Entities:  

Mesh:

Year:  1997        PMID: 9224455     DOI: 10.1016/s0161-6420(97)30183-3

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  17 in total

1.  Depth evaluation of intended vs actual intacs intrastromal ring segments using optical coherence tomography.

Authors:  R Barbara; A Barbara; M Naftali
Journal:  Eye (Lond)       Date:  2015-10-23       Impact factor: 3.775

2.  Tissue reaction after intrastromal corneal ring implantation in an experimental animal model.

Authors:  Lucía Ibares-Frías; Patricia Gallego; Roberto Cantalapiedra-Rodríguez; María Cruz Valsero; Santiago Mar; Jesús Merayo-Lloves; María Carmen Martínez-García
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-03-07       Impact factor: 3.117

3.  Intrastromal corneal ring segments (ICRs): three- and six months results.

Authors:  R H Wijdh; G van Rij
Journal:  Doc Ophthalmol       Date:  2000       Impact factor: 2.379

4.  New surgical approaches to the management of keratoconus and post-LASIK ectasia.

Authors:  Bryan U Tan; Tracy L Purcell; Luis F Torres; David J Schanzlin
Journal:  Trans Am Ophthalmol Soc       Date:  2006

5.  Intrastromal corneal ring segment implantation (keraring 355°) in patients with central keratoconus: 6-month follow-up.

Authors:  Khosrow Jadidi; Seyed Aliasghar Mosavi; Farhad Nejat; Mostafa Naderi; Leila Janani; Sara Serahati
Journal:  J Ophthalmol       Date:  2015-01-05       Impact factor: 1.909

Review 6.  A Comparison of Different Operating Systems for Femtosecond Lasers in Cataract Surgery.

Authors:  B M Wu; G P Williams; A Tan; J S Mehta
Journal:  J Ophthalmol       Date:  2015-09-21       Impact factor: 1.909

7.  Efficacy of intacs intrastromal corneal ring segment relative to depth of insertion evaluated with anterior segment optical coherence tomography.

Authors:  Hassan Hashemi; Alireza Yazdani-Abyaneh; Amirhushang Beheshtnejad; Mahmood Jabbarvand; Ahmad Kheirkhah; Seyed Reza Ghaffary
Journal:  Middle East Afr J Ophthalmol       Date:  2013 Jul-Sep

8.  Nine-year follow-up of intacs implantation for keratoconus.

Authors:  George D Kymionis; Michael A Grentzelos; Vasilios F Diakonis; Aristofanis I Pallikaris; Ioannis G Pallikaris
Journal:  Open Ophthalmol J       Date:  2009-12-08

9.  Morphological characterization and clinical effects of stromal alterations after intracorneal ring segment implantation in keratoconus.

Authors:  Loïc Hamon; Ursula Schlötzer-Schrehardt; Fidelis A Flockerzi; Berthold Seitz; Loay Daas
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-02-02       Impact factor: 3.535

10.  Safety and efficacy of Intacs in Indian eyes with keratoconus: an initial report.

Authors:  Rohit Shetty; Kannan M Narayana; Kurian Mathew; D Anand; Prerna Mhaske; Bhujang K Shetty
Journal:  Indian J Ophthalmol       Date:  2009 Mar-Apr       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.