R Barbara1, A Barbara2, M Naftali3.
Abstract
PurposeEvaluation of actual vs intended intrastromal corneal ring segments (ICRS) implantation depth as measured by anterior segment optical coherence tomography (OCT)MethodsProspective study evaluating 30 Intacs segments implanted manually in 19 eyes of 15 patients suffering from keratoconus. Segment depth evaluation was performed using anterior segment OCT. Measurements were performed above and below the segment at 3 points in relation to the incision site. Statistical analysis was performed using the SAS software for ANOVA, matched t-test, and GLIMMIX procedure.ResultsIntacs segment depth was 153-μm shallower than intended (58% vs 80%). Segment layout demonstrated the proximal and distal portions to be 13-μm shallower and 12-μm deeper (on average), respectively. Intacs segment thickness does not influence implantation depth. Intacs segments implanted in the same eye do not share similar implantation depths. Stromal compression is likely to occur.ConclusionIntacs are implanted at a shallower depth than intended. The 'pocketing' stage prior to implantation most likely has a stronger effect on the segment's final implantation depth than does the incisions' depth.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26493037 PMCID: PMC4709544 DOI: 10.1038/eye.2015.202
Source DB: PubMed Journal: Eye (Lond) ISSN: 0950-222X Impact factor: 3.775