Literature DB >> 9218301

Thrombotic, infectious, and procedural complications of the jugular bulb catheter in the intensive care unit.

W M Coplin1, G E O'Keefe, M S Grady, G A Grant, K S March, H R Winn, A M Lam.   

Abstract

OBJECTIVE: An assessment of the thrombotic, infectious, and technical complications of continuous jugular bulb catheter monitoring in the intensive care unit (ICU) was made.
METHODS: Over a 1-year period, 44 patients suffering from traumatic brain injury, subarachnoid hemorrhage, or stroke received jugular bulb catheter monitoring in the ICU. They were followed for catheter insertion complications and the development of bacteremia. In 20 patients chosen randomly, an ultrasonographic evaluation was performed after removal of the catheter for an assessment of internal jugular vein thrombosis.
RESULTS: Of the 44 patients, 1 became bacteremic; the source was identified as a thoracostomy site. Among the complications related to the 44 catheter insertions, there were 2 instances of carotid artery puncture (4.5%), 1 misplaced catheter (thoracic placement), and 1 clinically insignificant hematoma. Of the 20 patients investigated with ultrasonography, 8 (40%) had nonobstructive, subclinical internal jugular vein thrombi after jugular bulb catheter monitoring (95% confidence interval, 19-61%). The median monitoring duration was 3 days (range, 1-6 d). No clinical factor was identified to be associated with thrombus formation.
CONCLUSION: We conclude the following: 1) the risk of bacteremia related to the jugular bulb catheter was negligible; 2) complications related to catheter insertion were rare and clinically insignificant; and 3) the incidence of subclinical internal jugular vein thrombosis after jugular bulb catheter monitoring is considerable. Although it is worthy to note this complication, no patient with a thrombus became symptomatic in the present series. The risk-benefit assessment of this monitoring technique must include consideration of subclinical thrombosis.

Entities:  

Mesh:

Year:  1997        PMID: 9218301     DOI: 10.1097/00006123-199707000-00022

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  17 in total

1.  Circulatory effects of internal jugular vein compression: a computer simulation study.

Authors:  R Bosnjak; M Kordas
Journal:  Med Biol Eng Comput       Date:  2002-07       Impact factor: 2.602

Review 2.  Update on multimodality monitoring.

Authors:  Chad M Miller
Journal:  Curr Neurol Neurosci Rep       Date:  2012-08       Impact factor: 5.081

3.  MRI estimation of global brain oxygen consumption rate.

Authors:  Varsha Jain; Michael C Langham; Felix W Wehrli
Journal:  J Cereb Blood Flow Metab       Date:  2010-04-21       Impact factor: 6.200

4.  Simultaneous bedside assessment of global cerebral blood flow and effective cerebral perfusion pressure in patients with intracranial hypertension.

Authors:  M Jägersberg; C Schaller; J Boström; B Schatlo; M Kotowski; C Thees
Journal:  Neurocrit Care       Date:  2010-04       Impact factor: 3.210

Review 5.  MRI-based methods for quantification of the cerebral metabolic rate of oxygen.

Authors:  Zachary B Rodgers; John A Detre; Felix W Wehrli
Journal:  J Cereb Blood Flow Metab       Date:  2016-04-18       Impact factor: 6.200

6.  Comparison of tissue oximeters on a liquid phantom with adjustable optical properties.

Authors:  S Kleiser; N Nasseri; B Andresen; G Greisen; M Wolf
Journal:  Biomed Opt Express       Date:  2016-07-11       Impact factor: 3.732

7.  Comparison of tissue oximeters on a liquid phantom with adjustable optical properties: an extension.

Authors:  S Kleiser; D Ostojic; B Andresen; N Nasseri; H Isler; F Scholkmann; T Karen; G Greisen; M Wolf
Journal:  Biomed Opt Express       Date:  2017-12-05       Impact factor: 3.732

Review 8.  Multimodality monitoring in severe traumatic brain injury: the role of brain tissue oxygenation monitoring.

Authors:  Jamin M Mulvey; Nicholas W C Dorsch; Yugan Mudaliar; Erhard W Lang
Journal:  Neurocrit Care       Date:  2004       Impact factor: 3.210

9.  Accuracy of the cylinder approximation for susceptometric measurement of intravascular oxygen saturation.

Authors:  Cheng Li; Michael C Langham; Charles L Epstein; Jeremy F Magland; Jue Wu; James Gee; Felix W Wehrli
Journal:  Magn Reson Med       Date:  2011-08-19       Impact factor: 4.668

10.  Infected internal jugular vein thrombus in a case of infected arterio-venous fistula for dialysis access.

Authors:  Kaushik Mukherjee; Ushnish Chakrabart; Pinaki Mazumder; Madhusudan Pal; Kallol Dasbaksi; Plaban Mukherjee
Journal:  Ann Vasc Dis       Date:  2014-08-30
View more