DXA longitudinal quality control: a comparison of inbuilt quality assurance, visual inspection, multi-rule Shewhart charts and Cusum analysis.

S W Garland1, B Lees, J C Stevenson.   

Abstract

The performance of dual-energy X-ray absorptiometry (DXA) instruments can be monitored using various quality control (QC) procedures. It has not been established which of these is most appropriate. The aim of this study was to determine which of four QC procedures is the best to use for longitudinal monitoring. Eighteen centres with DXA instruments scanned an aluminium spine phantom weekly for up to 16 months, and the bone mineral density data were used for the QC procedures. The methods investigated were the instrument's inbuilt quality assurance (QA) procedure, visual inspection, multi-rule Shewhart control charts, and Cusum analysis using a truncated-V mask. True and false positive fractions (TPF and FPF) of each method were calculated, including those for a range of action levels for the Shewhart charts and dimensions of the Cusum mask. For Shewhart, the action levels giving the most desirable TPF and FPF were whole multiples of the standard deviation (SD). For Cusum, the most desirable mask dimensions were 3.6 SD for the total height of the vertical section and 0.9 SD per data point for the gradient of the wings. Predictive power of each method as a means of fault detection was decided by the number of faults detected out of a total of 8 non-mechanical faults subsequently diagnosed. The inbuilt QA detected 2, visual inspection 7, Shewhart chart 7 and Cusum analysis 7. The FPFs were: visual inspection 0.09, Shewhart 0.04, Cusum 0.08. At these levels of FPF, the average time in days (range) from onset of a fault to detection was 39 (6-82) for visual inspection, 39 (4-116) for Shewhart and 21 (1-49) for Cusum. All three "phantom" methods are excellent for DXA QC, with modified Cusum analysis being the most effective. The inbuilt QA appears of little use on its own for longitudinal QC.

Entities:  

Mesh:

Year:  1997        PMID: 9205636     DOI: 10.1007/bf01622294

Source DB:  PubMed          Journal:  Osteoporos Int        ISSN: 0937-941X            Impact factor:   4.507


  13 in total

1.  Basic principles of ROC analysis.

Authors:  C E Metz
Journal:  Semin Nucl Med       Date:  1978-10       Impact factor: 4.446

2.  Longitudinal precision of dual-energy x-ray absorptiometry in a multicenter study. The Nafarelin/Bone Study Group.

Authors:  E S Orwoll; S K Oviatt
Journal:  J Bone Miner Res       Date:  1991-02       Impact factor: 6.741

3.  An evaluation of cusum analysis and control charts applied to quantitative gamma-camera uniformity parameters for automated quality control.

Authors:  A C Knight; E D Williams
Journal:  Eur J Nucl Med       Date:  1992

4.  The use of control charts in the clinical laboratory.

Authors:  S LEVEY; E R JENNINGS
Journal:  Am J Clin Pathol       Date:  1950-11       Impact factor: 2.493

5.  Dual X-ray absorptiometry quality control: comparison of visual examination and process-control charts.

Authors:  Y Lu; A K Mathur; B A Blunt; C C Gluer; A S Will; T P Fuerst; M D Jergas; K N Andriano; S R Cummings; H K Genant
Journal:  J Bone Miner Res       Date:  1996-05       Impact factor: 6.741

6.  Performance characteristics of rules for internal quality control: probabilities for false rejection and error detection.

Authors:  J O Westgard; T Groth; T Aronsson; H Falk; C H de Verdier
Journal:  Clin Chem       Date:  1977-10       Impact factor: 8.327

7.  Factors influencing the quality of analytical methods. A systems analysis, with use of computer simulation.

Authors:  T Aronsson; C H de Verdier; T Groth
Journal:  Clin Chem       Date:  1974-07       Impact factor: 8.327

8.  Quality control of DXA instruments in multicenter trials.

Authors:  K G Faulkner; M R McClung
Journal:  Osteoporos Int       Date:  1995       Impact factor: 4.507

9.  A multi-rule Shewhart chart for quality control in clinical chemistry.

Authors:  J O Westgard; P L Barry; M R Hunt; T Groth
Journal:  Clin Chem       Date:  1981-03       Impact factor: 8.327

10.  Precision of dual-energy x-ray absorptiometry: development of quality control rules and their application in longitudinal studies.

Authors:  E S Orwoll; S K Oviatt; J A Biddle
Journal:  J Bone Miner Res       Date:  1993-06       Impact factor: 6.741

View more
  1 in total

Review 1.  Quality Control of DXA System and Precision Test of Radio-technologists.

Authors:  Ho-Sung Kim; Seoung-Oh Yang
Journal:  J Bone Metab       Date:  2014-02-28
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.