The effect of dose rate dependence of p-type silicon detectors on linac relative dosimetry.

D Wilkins1, X A Li, J Cygler, L Gerig.   

Abstract

Cumulative radiation damage to silicon semiconductor diode detectors can induce dose rate dependent sensitivity, a concern in the pulsed beam of a linac. Two p-Si diode photon detectors were used in this study, diodes A and B. Both were preirradiated by the supplier to 5 kGy, with diode A receiving an estimated 8 kGy from measurements, and diode B, 25 kGy. At 6 MV, the PDD measured with diode B was lower (by 4.4% at a depth of 25 cm) than diode A. Using SSD to vary the dose per pulse from 0.02 to 0.64 mGy/pulse, diode A was dose rate independent (within 2%), while the sensitivity of diode B changed by 13%. Silicon diode detectors should be checked regularly against ionization chambers in the pulsed beam of a linac, especially older high-resistivity diodes that have accumulated dose from high-energy photon beams.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9198022     DOI: 10.1118/1.597985

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Development of an applicator for eye lens dosimetry during radiotherapy.

Authors:  J M Park; J Lee; H S Kim; S-J Ye; J-I Kim
Journal:  Br J Radiol       Date:  2014-08-11       Impact factor: 3.039

2.  Characterization of a microSilicon diode detector for small-field photon beam dosimetry.

Authors:  Yuichi Akino; Masateru Fujiwara; Keita Okamura; Hiroya Shiomi; Hirokazu Mizuno; Fumiaki Isohashi; Osamu Suzuki; Yuji Seo; Keisuke Tamari; Kazuhiko Ogawa
Journal:  J Radiat Res       Date:  2020-05-22       Impact factor: 2.724

3.  The consistency of Fletcher-Suit applicator geometry and of the rectal probe's position in high dose rate brachytherapy treatment fraction of cervix carcinoma.

Authors:  Jenő Pálvölgyi
Journal:  J Contemp Brachytherapy       Date:  2009-10-08

4.  "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Authors:  Khalsa Al Shukaili; Stéphanie Corde; Marco Petasecca; Vladimir Pereveratylo; Michael Lerch; Michael Jackson; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-05-22       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.