Literature DB >> 9196468

The bionic glove: an electrical stimulator garment that provides controlled grasp and hand opening in quadriplegia.

A Prochazka1, M Gauthier, M Wieler, Z Kenwell.   

Abstract

OBJECTIVE: This report describes the operation of the Bionic Glove, a new functional electrical stimulation (FES) device designed to improve the function of the paralyzed hand after spinal cord injury (SCI) or stroke.
DESIGN: Signals from a sensor in the glove detecting voluntary wrist movement are used to control FES of muscles either to produce hand-grasp or to open the hand. When the glove is donned, conductive areas on its inside surface automatically make contact with self-adhesive electrodes on the skin. SETTING AND PATIENTS: This report concerns nine people with SCI who have used the device in their daily lives for up to a year or more. Measurements were made at clinics in Edmonton, Miami, and Chicago as part of a multicenter clinical trial. OUTCOME MEASURES AND
RESULTS: The mean peak force of tenodesis grasp in the nine subjects increased from 2.6N (passive) to 11.3N (glove active). Active force was significantly greater than passive grasp force even when muscles were fatigued after repetitive grasp-release cycles. Most manual tasks improved significantly with the use of the glove, as judged by the number of tasks completed in a minute or the subjects' qualitative ratings of task difficulty.
CONCLUSION: The Bionic Glove can provide significant improvement of hand function in people with C6-C7 SCI.

Entities:  

Mesh:

Year:  1997        PMID: 9196468     DOI: 10.1016/s0003-9993(97)90426-3

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  31 in total

1.  An artificial grasping evaluation system for the paralysed hand.

Authors:  M C de Castro; A Cliquet Júnior
Journal:  Med Biol Eng Comput       Date:  2000-05       Impact factor: 2.602

Review 2.  Neural prostheses.

Authors:  A Prochazka; V K Mushahwar; D B McCreery
Journal:  J Physiol       Date:  2001-05-15       Impact factor: 5.182

3.  A lithographically-patterned, elastic multi-electrode array for surface stimulation of the spinal cord.

Authors:  Kathleen W Meacham; Richard J Giuly; Liang Guo; Shawn Hochman; Stephen P DeWeerth
Journal:  Biomed Microdevices       Date:  2008-04       Impact factor: 2.838

4.  Restoring voluntary grasping function in individuals with incomplete chronic spinal cord injury: pilot study.

Authors:  Naaz Kapadia; Vera Zivanovic; Milos R Popovic
Journal:  Top Spinal Cord Inj Rehabil       Date:  2013

5.  Changes in corticospinal excitability evoked by common peroneal nerve stimulation depend on stimulation frequency.

Authors:  C S Mang; O Lagerquist; D F Collins
Journal:  Exp Brain Res       Date:  2010-03-09       Impact factor: 1.972

Review 6.  Sensory control of normal movement and of movement aided by neural prostheses.

Authors:  Arthur Prochazka
Journal:  J Anat       Date:  2015-06-05       Impact factor: 2.610

Review 7.  Neurophysiology and neural engineering: a review.

Authors:  Arthur Prochazka
Journal:  J Neurophysiol       Date:  2017-05-31       Impact factor: 2.714

Review 8.  A synthesis of best evidence for the restoration of upper-extremity function in people with tetraplegia.

Authors:  Sukhvinder Kalsi-Ryan; Mary C Verrier
Journal:  Physiother Can       Date:  2011-10-20       Impact factor: 1.037

9.  Command control for functional electrical stimulation hand grasp systems using miniature accelerometers and gyroscopes.

Authors:  K Y Tong; A F T Mak; W Y Ip
Journal:  Med Biol Eng Comput       Date:  2003-11       Impact factor: 2.602

10.  Stimulation stability and selectivity of chronically implanted multicontact nerve cuff electrodes in the human upper extremity.

Authors:  Katharine H Polasek; Harry A Hoyen; Michael W Keith; Robert F Kirsch; Dustin J Tyler
Journal:  IEEE Trans Neural Syst Rehabil Eng       Date:  2009-09-22       Impact factor: 3.802

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.