Proton pencil beams in water, in a format suitable for treatment planning algorithms and covering the radiotherapy energy range (50-250 MeV), have been calculated using a modified version of the Monte Carlo code PTRAN. A simple analytical model has also been developed for calculating proton broad-beam dose distributions which is in excellent agreement with the Monte Carlo calculations. Radial dose distributions are also calculated analytically and narrow proton pencil-beam dose distributions derived. The physical approximations in the Monte Carlo code and in the analytical model together with their limitations are discussed. Examples showing the use of the calculated set of proton pencil beams as input to an existing photon treatment planning algorithm based on biological optimization are given for fully 3D scanned proton pencil beams; these include intensity modulated beams with range shift and scanning in the transversal plane.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9194127     DOI: 10.1088/0031-9155/42/6/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Benchmark measurements and simulations of dose perturbations due to metallic spheres in proton beams.

Authors:  Wayne D Newhauser; Laura Rechner; Dragan Mirkovic; Pablo Yepes; Nicholas C Koch; Uwe Titt; Jonas D Fontenot; Rui Zhang
Journal:  Radiat Meas       Date:  2013-11-01       Impact factor: 1.898

2.  Quantification of proton dose calculation accuracy in the lung.

Authors:  Clemens Grassberger; Juliane Daartz; Stephen Dowdell; Thomas Ruggieri; Greg Sharp; Harald Paganetti
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-04-11       Impact factor: 7.038

3.  Proton range verification in inhomogeneous tissue: Treatment planning system vs. measurement vs. Monte Carlo simulation.

Authors:  Dae-Hyun Kim; Sungkoo Cho; Kwanghyun Jo; EunHyuk Shin; Chae-Seon Hong; Youngyih Han; Tae-Suk Suh; Do Hoon Lim; Doo Ho Choi
Journal:  PLoS One       Date:  2018-03-05       Impact factor: 3.240

4.  Proton Therapy for Mandibula Plate Phantom.

Authors:  Güler Burcu Senirkentli; Fatih Ekinci; Erkan Bostanci; Mehmet Serdar Güzel; Özlem Dağli; Ahmed M Karim; Alok Mishra
Journal:  Healthcare (Basel)       Date:  2021-02-04

5.  A fast Monte Carlo code for proton transport in radiation therapy based on MCNPX.

Authors:  Keyvan Jabbari; Jan Seuntjens
Journal:  J Med Phys       Date:  2014-07
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.