Pulse oximeter performance during desaturation and resaturation: a comparison of seven models.

N S Trivedi1, A F Ghouri, E Lai, N K Shah, S J Barker.   

Abstract

STUDY
OBJECTIVE: To compare pulse oximeter performance during induced hypoxemia.
DESIGN: Prospective investigation in human volunteers.
SETTING: Laboratory facility at a university medical center. PATIENTS: 8 unanesthetized, healthy ASA physical status I volunteers.
INTERVENTIONS: We evaluated the accuracy and response times of seven popular pulse oximeters during induced hypoxemia. Arterial blood fractional oxygen saturation (SaO2) measurements were performed simultaneously and considered a gold standard.
MEASUREMENTS AND MAIN RESULTS: All oximeters were accurate (+/-2%) while subjects were breathing room air. During maximal hypoxemia (induced by breathing a FIO2 = 10% in nitrogen), large differences were noted between oxygen saturation as measured by pulse oximetry (SpO2) and SaO2 values, with pulse oximeters consistently underreporting SpO2 when actual SaO2 values were 75% or less. The Ohmeda 3740 (Ohmeda, Boulder, CO) using an ear probe was the first to detect desaturation (change in SpO2 > 3%) in 4 of 8 subjects (p < 0.05), and the Nellcor N200 reflectance oximeter (Nellcor, Inc., Pleasanton, CA) was first in 3 of 8 subjects (p < 0.05). During resaturation (after administering 100% oxygen), the Novametrix Oxypleth (Novametrix, Wallingford, CT) was significantly faster than other oximeters (p < 0.05) to return to baseline (SpO2 = 98%).
CONCLUSION: Most models of oximeters tested performed well when hemoglobin oxygen saturation was high, but all were inaccurate when SaO2 was approximately 75%. During induced hypoxemia, there were significant differences in the response times of oximeters tested, with no model demonstrably superior to others in all measures of performance.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9172023     DOI: 10.1016/s0952-8180(97)00037-8

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  9 in total

1.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

2.  Oximetry feedback flow control simulation for oxygen therapy.

Authors:  Mario G Iobbi; Anita K Simonds; Robert J Dickinson
Journal:  J Clin Monit Comput       Date:  2007-01-05       Impact factor: 2.502

3.  Significant Delay in the Detection of Desaturation between Finger Transmittance and Earlobe Reflectance Oximetry Probes during Fiberoptic Bronchoscopy: Analysis of 104 Cases.

Authors:  Barak Pertzov; Elitsur Brachfeld; Avraham Unterman; Evgeni Gershman; Nader Abdel-Rahman; Dror Rosengarten; Mordechai Reuven Kramer
Journal:  Lung       Date:  2018-11-24       Impact factor: 2.584

4.  A comparison of response time to desaturation between tracheal oximetry and peripheral oximetry.

Authors:  Li Wang; Wei Wei; Ming Gong; Ling Mu
Journal:  J Clin Monit Comput       Date:  2010-03-12       Impact factor: 2.502

5.  Effect of concurrent oxygen therapy on accuracy of forecasting imminent postoperative desaturation.

Authors:  Hisham ElMoaqet; Dawn M Tilbury; Satya Krishna Ramachandran
Journal:  J Clin Monit Comput       Date:  2014-10-19       Impact factor: 2.502

6.  Pulse Oximeter Performance during Rapid Desaturation.

Authors:  Lenka Horakova; Karel Roubik
Journal:  Sensors (Basel)       Date:  2022-06-02       Impact factor: 3.847

7.  Agreement in computer-assisted manual scoring of polysomnograms across sleep centers.

Authors:  Samuel T Kuna; Ruth Benca; Clete A Kushida; James Walsh; Magdy Younes; Bethany Staley; Alexandra Hanlon; Allan I Pack; Grace W Pien; Atul Malhotra
Journal:  Sleep       Date:  2013-04-01       Impact factor: 5.849

8.  Increased PIO2 at Exhaustion in Hypoxia Enhances Muscle Activation and Swiftly Relieves Fatigue: A Placebo or a PIO2 Dependent Effect?

Authors:  Rafael Torres-Peralta; José Losa-Reyna; David Morales-Alamo; Miriam González-Izal; Ismael Pérez-Suárez; Jesús G Ponce-González; Mikel Izquierdo; José A L Calbet
Journal:  Front Physiol       Date:  2016-08-17       Impact factor: 4.566

9.  Prior oxygenation, but not chemoreflex responsiveness, determines breath-hold duration during voluntary apnea.

Authors:  Christina D Bruce; Emily R Vanden Berg; Jamie R Pfoh; Craig D Steinback; Trevor A Day
Journal:  Physiol Rep       Date:  2021-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.