Quantitative LVEF and qualitative regional function from gated thallium-201 perfusion SPECT.

G Germano1, J Erel, H Kiat, P B Kavanagh, D S Berman.   

Abstract

UNLABELLED: This study investigates the feasibility of routine clinical 201Tl gated perfusion SPECT (gated Tl), and compares quantitative left ventricular ejection fraction (LVEF) and visually-assessed regional wall motion and thickening to analogous values obtained from 99mTc-sestamibi gated perfusion SPECT (gated MIBI).
METHODS: We studied 121 patients with a rest gated Tl (3-3.5 mCi, 35 sec/ projection/poststress gated MIBI (25-30 mCi, 25 sec/projection) separate dual-isotope protocol on a 90 degrees dual-detector camera. Automatic quantitation of LVEFs was accomplished using previously developed and validated software, while visual scoring of motion and thickening was performed using four-point scales.
RESULTS: Average myocardial counts were lower in gated Tl images (306 +/- 81 counts/pixel) compared to gated MIBI images (789 +/- 237 counts/pixel). The quality of gated Tl images was ranked as excellent, good, fair and poor in 24.0%, 42.1%, 24.8% and 9.1%, respectively, of the patients, compared to 43.0%, 43.8%, 9.1% and 4.1%, respectively, for gated MIBI images. Quantitative-gated Tl and gated MIBI LVEFs correlated well (y = 0.11 + 1.05x, r = 0.918, SEE = 6.35). Possible poststress myocardial stunning may have caused gated Tl LVEFs to overestimate gated MIBI LVEFs by a larger (p = 0.03) amount in ischemic patients (n = 47, y = -0.69 + 1.09x, r = 0.914, s.e.e. = 6.44) compared to nonischemic patients (n = 64, y = -1.58 + 1.05x, r = 0.919, s.e.e. = 5.93), the residual difference in LVEFs for this latter group being likely due to different isotope resolution in conjunction with small left ventricles. Exact agreement between gated Tl and gated MIBI segmental myocardial function in 41 nonischemic patients was 92.2% (kappa = 0.619) and 95.4% (kappa = 0.586) for motion and thickening scores, respectively.
CONCLUSION: Thallium-201 gated SPECT imaging can be effectively performed on the majority of patients in our clinical environment and offers the opportunity to assess both myocardial perfusion and function using one injection and one imaging sequence, similarly to what is done with 99mTc-based agents.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9170440

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  54 in total

1.  Segmentation of gated Tl-SPECT images and computation of ejection fraction: a different approach.

Authors:  P Brigger; S L Bacharach; G Srinivasan; K A Nour; J A Carson; V Dilsizian; A Aldroubi; M Unser
Journal:  J Nucl Cardiol       Date:  1999 May-Jun       Impact factor: 5.952

2.  Proceedings of the 4th Invitational Wintergreen Conference. Wintergreen, Virginia, USA. July 12-14, 1998. Abstracts.

Authors: 
Journal:  J Nucl Cardiol       Date:  1999 Jan-Feb       Impact factor: 5.952

3.  Evaluation of left ventricular wall motion, volumes, and ejection fraction by gated myocardial tomography with technetium 99m-labeled tetrofosmin: a comparison with cine magnetic resonance imaging.

Authors:  P Vaduganathan; Z X He; G W Vick; J J Mahmarian; M S Verani
Journal:  J Nucl Cardiol       Date:  1999 Jan-Feb       Impact factor: 5.952

4.  Assessment of left ventricular diastolic function with electrocardiography-gated myocardial perfusion SPECT: comparison with multigated equilibrium radionuclide angiography.

Authors:  S Kumita; K Cho; H Nakajo; M Toba; M Uwamori; S Mizumura; T Kumazaki; J Sano; S Sakai; K Munakata
Journal:  J Nucl Cardiol       Date:  2001 Sep-Oct       Impact factor: 5.952

5.  Comparison of two three-dimensional gated SPECT methods with thallium in patients with large myocardial infarction.

Authors:  P Véra; R Koning; A Cribier; A Manrique
Journal:  J Nucl Cardiol       Date:  2000 Jul-Aug       Impact factor: 5.952

6.  Comparative value of ECG-gated blood pool SPET and ECG-gated myocardial perfusion SPET in the assessment of global systolic left ventricular function.

Authors:  Doumit Daou; Didier Vilain; Patrice Colin; Rachida Lebtahi; Thierry Fourme; Carlos Coaguila; Abdel Benada; Ilana Idy-Peretti; Michel Slama; Dominique Le Guludec
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-04-04       Impact factor: 9.236

Review 7.  Measurement of ventricular function with radionuclide techniques.

Authors:  Kim A Williams
Journal:  Curr Cardiol Rep       Date:  2003-01       Impact factor: 2.931

8.  Worsening of left ventricular ejection fraction induced by dipyridamole on Tl-201 gated myocardial perfusion imaging predicts significant coronary artery disease.

Authors:  Guang-Uei Hung; Kung-Wei Lee; Ching-Pei Chen; Kuang-Tao Yang; Wan-Yu Lin
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

9.  Evaluation of an attenuation correction method for thallium-201 myocardial perfusion tomographic imaging of patients with low likelihood of coronary artery disease.

Authors:  P Chouraqui; S Livschitz; T Sharir; N Wainer; M Wilk; I Moalem; J Baron
Journal:  J Nucl Cardiol       Date:  1998 Jul-Aug       Impact factor: 5.952

10.  Evaluation of immediate post-stress wall motion on adenosine stress/rest thallium-201 gated myocardial SPECT.

Authors:  Won Woo Lee; Eun Kyung Park; Jae Seon Eo; Sang Woo Lee; Cheol Ho Kim; Young So; Dong Soo Lee; June-Key Chung; Myung Chul Lee; Sang Eun Kim