Literature DB >> 9169713

Evaluation of an on-line patient exposure meter in neuroradiology.

N A Gkanatsios1, W Huda, K R Peters, J A Freeman.   

Abstract

PURPOSE: To assess the clinical performance and usefulness of an on-line patient exposure meter installed on a neuroradiologic biplane imaging system.
MATERIALS AND METHODS: A commercial on-line patient exposure meter was installed on each plane of a biplane neuroradiologic imaging system. The meter computed skin exposures on the basis of selected technique factors (tube potential and current) and information about patient location relative to the x-ray tube. Simulations were performed to measure the system accuracy with an angiographic anthropomorphic head phantom with the skin exposures measured with an ionization chamber. Skin doses were computed for 114 consecutive patients who underwent diagnostic and interventional neuroradiologic procedures.
RESULTS: Agreement between measured and computed skin exposures in fluoroscopy, plain radiography, and digital imaging was generally within 5% of the true skin dose. For all fluoroscopic and radiographic procedures, total median skin doses were 1.20 and 0.64 Gy for the frontal and lateral planes, respectively. In both planes, patient skin doses resulted primarily from digital subtraction angiographic acquisitions. In 29 (25%) patients, the skin dose exceeded 2.00 Gy, but no radiation-induced deterministic effects were observed.
CONCLUSION: An on-line patient exposure meter can provide accurate radiation skin dose data in patients undergoing diagnostic and therapeutic neuroradiologic procedures.

Entities:  

Mesh:

Year:  1997        PMID: 9169713     DOI: 10.1148/radiology.203.3.9169713

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  7 in total

1.  Skin dose mapping for fluoroscopically guided interventions.

Authors:  Perry B Johnson; David Borrego; Stephen Balter; Kevin Johnson; Daniel Siragusa; Wesley E Bolch
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

2.  Dose measurement on both patients and operators during neurointerventional procedures using photoluminescence glass dosimeters.

Authors:  T Moritake; Y Matsumaru; T Takigawa; K Nishizawa; A Matsumura; K Tsuboi
Journal:  AJNR Am J Neuroradiol       Date:  2008-08-21       Impact factor: 3.825

3.  Fluoroscopy fade for embolization of vein of Galen malformation.

Authors:  Curtis A Given; Ike C Thacker; Michael D Baker; P Pearse Morris
Journal:  AJNR Am J Neuroradiol       Date:  2003-02       Impact factor: 3.825

4.  Patient skin dose during neuroembolization by multiple-point measurement using a radiosensitive indicator.

Authors:  S Suzuki; S Furui; Y Matsumaru; S Nobuyuki; M Ebara; T Abe; D Itoh
Journal:  AJNR Am J Neuroradiol       Date:  2008-04-03       Impact factor: 3.825

5.  Endovascular image-guided interventions (EIGIs).

Authors:  Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

Review 6.  Patient dose reduction during voiding cystourethrography.

Authors:  Valerie L Ward
Journal:  Pediatr Radiol       Date:  2006-09

Review 7.  Methods for measuring fluoroscopic skin dose.

Authors:  Stephen Balter
Journal:  Pediatr Radiol       Date:  2006-09
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.