Literature DB >> 9151186

Stress fracture of the femoral component in total knee replacement: a report of 3 cases.

M Wada1, S Imura, A Bo, H Baba, T Miyazaki.   

Abstract

We report 3 cases of stress fracture of the femoral component (small size Whiteside Ortholoc II) in total knee replacements which occurred at 32, 52, and 73 months after operation. The site was at the junction between the medial posterior bevelled surface and the posterior flange. The most likely cause of failure is due to the thinness of the metal at this point.

Entities:  

Mesh:

Year:  1997        PMID: 9151186      PMCID: PMC3615681          DOI: 10.1007/s002640050118

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  5 in total

1.  A rare cause for knee pain: fracture of the femoral component after TKR. A case report.

Authors:  C Luring; L Perlick; T Schubert; M Tingart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-10-06       Impact factor: 4.342

2.  Fracture at the stem-condylar junction of a modular femoral prosthesis in a varus-valgus constrained total knee arthroplasty.

Authors:  D D Nikolopoulos; I G Polyzois; E A Magnissalis; P F Bernard; I V Michos
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-08-25       Impact factor: 4.342

Review 3.  [Alternative materials and solutions in total knee arthroplasty for patients with metal allergy].

Authors:  R Bader; P Bergschmidt; A Fritsche; S Ansorge; P Thomas; W Mittelmeier
Journal:  Orthopade       Date:  2008-02       Impact factor: 1.087

4.  Tibial baseplate fracture associated with polyethylene wear and osteolysis after total knee arthroplasty.

Authors:  Deniz Cankaya; Serdar Yilmaz; Alper Deveci; Bulent Ozkurt
Journal:  Int J Surg Case Rep       Date:  2013-12-21

5.  FRACTURE OF THE TIBIAL COMPONENT IN TOTAL KNEE ARTHROPLASTY: REPORT ON TWO CASES.

Authors:  Idemar Monteiro da Palma; Rodrigo Pires E Albuquerque; João Maurício Barretto
Journal:  Rev Bras Ortop       Date:  2015-12-08
  5 in total