Literature DB >> 21866350

Fracture at the stem-condylar junction of a modular femoral prosthesis in a varus-valgus constrained total knee arthroplasty.

D D Nikolopoulos1, I G Polyzois, E A Magnissalis, P F Bernard, I V Michos.   

Abstract

UNLABELLED: We present a rare case of fracture of the modular femoral component in a varus-valgus constrained total knee arthroplasty (TKA) prosthesis which occurred 2 years after implantation. Both femoral and tibial components underwent a second revision using a constrained hinged TKA. Laboratory analysis revealed insufficient metaphyseal bone stock in association with inadequate cement mantle incorporation which made implant fixation solely reliant on the intramedullary stem construct. It was therefore hypothesized that high repetitive stresses over the junction of the modular component led to the fracture. LEVEL OF EVIDENCE: Case report, Level IV.

Entities:  

Mesh:

Year:  2011        PMID: 21866350     DOI: 10.1007/s00167-011-1652-4

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  17 in total

1.  Failure of the stem-condyle junction of a modular femoral stem in revision total knee arthroplasty: a report of five cases.

Authors:  L A Lim; R T Trousdale; D J Berry; A D Hanssen
Journal:  J Arthroplasty       Date:  2001-01       Impact factor: 4.757

2.  Acute locking in revision total knee arthroplasty due to disengagement of the locking screw.

Authors:  J B Nadkarni; D G Carden
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2004-10-09       Impact factor: 4.342

3.  Failure at the taper lock of a modular stemmed femoral implant in revision knee arthroplasty. A report of two cases and a retrieval analysis.

Authors:  Paul S Issack; Jocelyn M Cottrell; Samuel Delgado; Timothy M Wright; Thomas P Sculco; Edwin P Su
Journal:  J Bone Joint Surg Am       Date:  2007-10       Impact factor: 5.284

4.  Detection of locking bolt loosening in the stem-condyle junction of a modular femoral stem in revision total knee arthroplasty.

Authors:  Jae-Min Ahn; Jeung-Tak Suh
Journal:  J Arthroplasty       Date:  2009-08-13       Impact factor: 4.757

5.  Disengagement of a locking screw from a modular stem in revision total knee arthroplasty. A report of three cases.

Authors:  G H Westrich; C Hidaka; R E Windsor
Journal:  J Bone Joint Surg Am       Date:  1997-02       Impact factor: 5.284

6.  Fracture of femoral component in a resurfacing total knee arthroplasty.

Authors:  John Michos; John Rallis; Apostolos Fassoulas
Journal:  J Arthroplasty       Date:  2006-06-21       Impact factor: 4.757

7.  Fracture of the femoral component in cementless total knee arthroplasty.

Authors:  L A Whiteside; D R Fosco; J G Brooks
Journal:  Clin Orthop Relat Res       Date:  1993-01       Impact factor: 4.176

8.  Breakage of a Total Condylar III knee prosthesis. A case report.

Authors:  E J McPherson; K G Vince
Journal:  J Arthroplasty       Date:  1993-10       Impact factor: 4.757

9.  Early results of the use of tantalum femoral cones for revision total knee arthroplasty.

Authors:  James L Howard; Jeremy Kudera; David G Lewallen; Arlen D Hanssen
Journal:  J Bone Joint Surg Am       Date:  2011-03-02       Impact factor: 5.284

10.  Femoral component fracture due to osteolysis after cemented mobile-bearing total knee arthroplasty.

Authors:  Chang Dong Han; Chang Wook Han; Ik Hwan Yang
Journal:  J Arthroplasty       Date:  2008-04-25       Impact factor: 4.757

View more
  6 in total

1.  Total and intercondylar notch bone resection in posterior stabilized knee arthroplasty: analysis of five manufacturer designs.

Authors:  Murat Bozkurt; Mesut Tahta; Safa Gursoy; Mustafa Akkaya
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-11-14       Impact factor: 4.342

2.  Ten-Year Results of Primary and Revision Condylar-Constrained Total Knee Arthroplasty in Patients with Severe Coronal Plane Instability.

Authors:  Andrea Camera; Stefano Biggi; Gabriele Cattaneo; Giovanni Brusaferri
Journal:  Open Orthop J       Date:  2015-08-31

3.  A clinical study of the rotational alignment of the femoral component in total knee arthroplasty.

Authors:  Liangjia Ding; Xiaomin Liu; Changlu Liu; Yingli Liu
Journal:  J Phys Ther Sci       Date:  2015-07-22

4.  Optimum stem length for mitigation of periprosthetic fracture risk following primary total knee arthroplasty: a finite element study.

Authors:  Noel Conlisk; Colin R Howie; Pankaj Pankaj
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-11-03       Impact factor: 4.342

5.  Trunnion Failure in Revision Total Knee Arthroplasty.

Authors:  Samir Chihab; Bryan M Grommersch; Greg A Erens; George N Guild
Journal:  Arthroplast Today       Date:  2021-02-26

6.  Failure at the femoral stem extension-condylar interface in a rotating hinge knee without radiographic evidence of loosening.

Authors:  Michael Kahan; Christopher Estes
Journal:  Arthroplast Today       Date:  2019-12-12
  6 in total