Literature DB >> 9146519

An in vitro comparison of the mechanical characteristics of three sagittal ramus osteotomy fixation techniques.

M T Murphy1, R H Haug, J E Barber.   

Abstract

PURPOSE: The purpose of this in vitro investigation was to compare the mechanical characteristics of three techniques for the rigid internal fixation of simulated sagittal ramus osteotomies in terms of their ability to resist vertical loads resembling masticatory forces.
MATERIALS AND METHODS: Fifteen bovine ribs of similar size and shape were subjected to uniform osteotomies resembling a sagittal ramus osteotomy. They were divided into three groups and fixed with three 2.0-mm bicortical positional screws in an inverted L-pattern, eight monocortical screws and a sagittal strut plate, or a hybrid system using three 2.0-mm bicortical positional screws, five 2.0-mm monocortical screws, and a sagittal strut plate. The repaired ribs were secured in a jig, subjected to vertically deforming forces, and evaluated for yield load, yield point, maximum load, displacement at maximum load, load at 3.0 mm displacement, and stiffness. The various groups were compared statistically with a Scheffe multiple comparison test.
RESULTS: No statistically significant differences existed between the positional screw group and the hybrid group (P > .05). No statistically significant differences were noted in yield displacement, yield load, displacement at maximum load, for any of the groups (P > .05). There were differences noted between the monocortical strut group and other groups in maximum load, load at 3.0 mm displacement, and stiffness (P > .05).
CONCLUSIONS: The monocortically fixed sagittal strut plate offers less resistance to maximum load and loads at 3.0 mm displacement and is less stiff than either a three-positional screw system or a hybrid system. No differences existed between the three positional screw systems and the hybrid system. However, given the low yield loads and yield displacements, these differences may not be clinically important.

Entities:  

Mesh:

Year:  1997        PMID: 9146519     DOI: 10.1016/s0278-2391(97)90699-6

Source DB:  PubMed          Journal:  J Oral Maxillofac Surg        ISSN: 0278-2391            Impact factor:   1.895


  9 in total

1.  Analysis of Mandibular Test Specimens Used to Assess a Bone Fixation System.

Authors:  Leandro Stocco Baccarin; Renato Correa Viana Casarin; Jorge Vicente Lopes-da-Silva; Luis Augusto Passeri
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2014-11-20

2.  In vitro biomechanical evaluation of the effect of an additional L-shaped plate on straight or box plate fixation in sagittal split ramus osteotomy using a bioabsorbable plate system.

Authors:  Kazuhiro Matsushita; Nobuo Inoue; Yasunori Totsuka
Journal:  Oral Maxillofac Surg       Date:  2011-03-04

3.  Mechanical and photoelastic analysis of conventional screws and cannulated screws for sagittal split osteotomy fixation: a comparative study.

Authors:  Cristina Jardelino de Lima; Saulo Gabriel Moreira Falci; Danillo Costa Rodrigues; Érica Cristina Marchiori; Roger Willian Fernandes Moreira
Journal:  Oral Maxillofac Surg       Date:  2015-06-06

4.  Mandibular Angle Fractures: A Clinical and Biomechanical Comparison-the Works of Ellis and Haug.

Authors:  Richard H Haug; Bethany L Serafin
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2008-11

5.  Comparison of the Stability of Mandibular Sagittal Osteotomy Fixation between Two Types of Titanium Miniplates: A Biomechanical Study in Sheep Mandibles.

Authors:  Guilherme Dos Santos Trento; Felipe Andrade Pires; Fábio André Dos Santos; Delson João da Costa; Nelson Luis Barbosa Rebellato; Leandro Eduardo Klüppel
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2017-05-18

6.  Comparison of strengths of five internal fixation methods used after bilateral sagittal split ramus osteotomy: An in vitro study.

Authors:  Farzin Sarkarat; Atiye Ahmady; Farzam Farahmand; Ali Fateh; Roozbeh Kahali; Amir Nourani; Vahid Rakhshan
Journal:  Dent Res J (Isfahan)       Date:  2020-08-14

Review 7.  TMJ response to mandibular advancement surgery: an overview of risk factors.

Authors:  José Valladares-Neto; Lucia Helena Cevidanes; Wesley Cabral Rocha; Guilherme de Araújo Almeida; João Batista de Paiva; José Rino-Neto
Journal:  J Appl Oral Sci       Date:  2014 Jan-Feb       Impact factor: 2.698

8.  Biomechanical in vitro evaluation of three stable internal fixation techniques used in sagittal osteotomy of the mandibular ramus: a study in sheep mandibles.

Authors:  Leandro Benetti de Olivera; Eduardo Sant' Ana; Antonio José Manzato; Fábio Luis Bunemer Guerra; G William Arnett
Journal:  J Appl Oral Sci       Date:  2012 Jul-Aug       Impact factor: 2.698

Review 9.  Evaluation of bite force after open reduction and internal fixation using microplates.

Authors:  S Tharani Kumar; Saurabh Saraf; S Prasanna Devi
Journal:  J Dent (Tehran)       Date:  2013-09-30
  9 in total