Literature DB >> 9124913

Obstruction of mechanical mitral prostheses: analysis of pathologic findings.

N Vitale1, A Renzulli, L Agozzino, A Pollice, N Tedesco, L de Luca Tupputi Schinosa, M Cotrufo.   

Abstract

BACKGROUND: The pathologic and echocardiographic findings observed in 87 patients with mitral valve obstruction were reviewed to ascertain the incidence of pannus formation versus that of thrombosis, the relationship between the two, and the time to the occurrence of pannus versus the time to thrombosis.
METHODS: Pannus morphology (concentric or eccentric), its location on the valve (atrial, ventricular, atrioventricular), and the presence and relationship of associated thrombi (atrial, ventricular, atrioventricular) were analyzed. The times between valve replacement and the occurrence of obstruction were also compared.
RESULTS: There were 10 caged-ball valves, 65 tilting-disc valves, and 12 bileaflet valves. Seventy-two patients underwent prosthetic replacement, and 15 underwent thrombolysis. Pannus alone was found in 27, pannus and thrombus in 39, and thrombus alone in 21. Primary thrombosis occurred earlier than pannus formation (p = 0.04); this was true for patients with bileaflet valves (p = 0.006) and those with tilting-disc valves (p = 0.04). Pannus was atrial in 19.7% (13/66), ventricular in 21.2% (14/66), and atrioventricular in 59.1% (39/66). Pannus morphology was concentric in 22.7% (15/66) and eccentric in 77.3% (51/66). Atrial secondary thrombi occur more often in patients with atrioventricular pannus (p = 0.04). Eight patients had reobstruction; this was caused by pannus formation in 5 and by thrombosis in 3. Five underwent reoperation, and 3 underwent thrombolysis. Reobstruction occurred earlier than the first event.
CONCLUSIONS: The frequency of pannus formation is much higher than that of thrombus formation, but thrombosis is of earlier onset than pannus formation. Thrombosis is due to the deposition of clots on the prosthesis, and a pannus occurs as the result of an inflammatory reaction developing on both valve surfaces.

Entities:  

Mesh:

Year:  1997        PMID: 9124913     DOI: 10.1016/s0003-4975(96)01391-4

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  17 in total

1.  Dysfunction of mechanical heart valve prosthesis: experience with surgical management in 48 patients.

Authors:  Wei-Guo Ma; Bin Hou; Adiljan Abdurusul; Ding-Xu Gong; Yue Tang; Qian Chang; Jian-Ping Xu; Han-Song Sun
Journal:  J Thorac Dis       Date:  2015-12       Impact factor: 2.895

2.  Limitations of multimodality imaging in the diagnosis of pannus formation in prosthetic aortic valve and review of the literature.

Authors:  Juan Bautista Soumoulou; Tomás Francisco Cianciulli; Andrea Zappi; Alberto Cozzarin; María Cristina Saccheri; Jorge Alberto Lax; Robert Guidoin; Ze Zhang
Journal:  World J Cardiol       Date:  2015-04-26

3.  Obstruction of the left ventricular outflow tract due to pannus formation after the implantation of a carbomedics aortic valve.

Authors:  Ken-ichi Imasaka; Masahiro Oe; Hironori Baba; Hitoshi Sumida
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2006-07

4.  Mitral valve replacement in patients younger than 6 years of age.

Authors:  T Katogi; R Aeba; Y Cho; Y Inoue; A Mitsumaru; S Takeuchi; S Kawada
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  1999-02

5.  St. Jude Medical prosthetic aortic valve malfunction due to pannus formation.

Authors:  Y Naito; M Hachida; T Shimabukuro; M Nonoyama; M Endo; H Koyanagi
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2000-11

Review 6.  Diagnostic evaluation of left-sided prosthetic heart valve dysfunction.

Authors:  Jesse Habets; Ricardo P Budde; Petr Symersky; Renee B van den Brink; Bas A de Mol; Willem P Mali; Lex A van Herwerden; Steven A Chamuleau
Journal:  Nat Rev Cardiol       Date:  2011-05-17       Impact factor: 32.419

Review 7.  Use of multidetector-row computed tomography scan to detect pannus formation in prosthetic mechanical aortic valves.

Authors:  Mohamed A Aladmawi; Claudio Pragliola; Olga Vriz; Domenico Galzerano
Journal:  J Thorac Dis       Date:  2017-04       Impact factor: 2.895

Review 8.  Mechanical valve obstruction: Review of diagnostic and treatment strategies.

Authors:  Jason Salamon; Jerson Munoz-Mendoza; Jared J Liebelt; Cynthia C Taub
Journal:  World J Cardiol       Date:  2015-12-26

9.  Subvalvular Pannus Overgrowth after Mosaic Bioprosthesis Implantation in the Aortic Position.

Authors:  Masanori Hirota; Tadashi Isomura; Minoru Yoshida; Chieko Katsumata; Fusahiko Ito; Masazumi Watanabe
Journal:  Ann Thorac Cardiovasc Surg       Date:  2015-12-02       Impact factor: 1.520

10.  Pannus overgrowth after mitral valve replacement with a Carpentier-Edwards pericardial bioprosthesis.

Authors:  Takeshi Oda; Seiya Kato; Eiki Tayama; Shuji Fukunaga; Hidetoshi Akashi; Shigeaki Aoyagi
Journal:  J Artif Organs       Date:  2009-03-29       Impact factor: 1.731

View more