Literature DB >> 16898646

Obstruction of the left ventricular outflow tract due to pannus formation after the implantation of a carbomedics aortic valve.

Ken-ichi Imasaka1, Masahiro Oe, Hironori Baba, Hitoshi Sumida.   

Abstract

Excessive pannus formation after implantation of a prosthetic valve is an infrequent but serious complication. A 69-year-old woman who had received a 19-mm CarboMedics aortic valve 11 years ago was readmitted to our hospital with dyspnea and chest oppression. Cineradiography did not show the restriction of valve movement. The aortic peak pressure gradient was calculated by Doppler echocardiography to be 104 mmHg. Based on the diagnosis of stenosis of the left ventricular outflow tract, the patient underwent reoperation. At reoperation, the pannus had formed circumferentially without disturbing the movement of the leaflet. A 19-mm St. Jude Medical Regent valve was implanted after enlargement of the aortic annulus. The patient's postoperative course was uneventful. We report this characteristic finding of pannus formation after the implantation of a CarboMedics valve in the aortic position.

Entities:  

Mesh:

Year:  2006        PMID: 16898646     DOI: 10.1007/pl00022259

Source DB:  PubMed          Journal:  Jpn J Thorac Cardiovasc Surg        ISSN: 1344-4964


  10 in total

1.  Obstruction of St Jude Medical valves in the aortic position: histology and immunohistochemistry of pannus.

Authors:  Hideki Teshima; Nobuhiko Hayashida; Hirohisa Yano; Masaru Nishimi; Eiki Tayama; Shuji Fukunaga; Hidetoshi Akashi; Takemi Kawara; Shigeaki Aoyagi
Journal:  J Thorac Cardiovasc Surg       Date:  2003-08       Impact factor: 5.209

2.  Acute thrombosis of prosthetic valves: a multivariate analysis of the risk factors for a lifethreatening event.

Authors:  A Renzulli; L De Luca; A Caruso; R Verde; D Galzerano; M Cotrufo
Journal:  Eur J Cardiothorac Surg       Date:  1992       Impact factor: 4.191

3.  The Björk-Shiley aortic prosthesis: flow characteristics, thrombus formation and tissue overgrowth.

Authors:  A P Yoganathan; W H Corcoran; E C Harrison; J R Carl
Journal:  Circulation       Date:  1978-07       Impact factor: 29.690

4.  St. Jude Medical prosthetic aortic valve malfunction due to pannus formation.

Authors:  Y Naito; M Hachida; T Shimabukuro; M Nonoyama; M Endo; H Koyanagi
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2000-11

5.  Obstruction of St Jude Medical valves in the aortic position: significance of a combination of cineradiography and echocardiography.

Authors:  S Aoyagi; M Nishimi; H Kawano; E Tayama; S Fukunaga; N Hayashida; H Akashi; T Kawara
Journal:  J Thorac Cardiovasc Surg       Date:  2000-07       Impact factor: 5.209

6.  Obstruction of mechanical mitral prostheses: analysis of pathologic findings.

Authors:  N Vitale; A Renzulli; L Agozzino; A Pollice; N Tedesco; L de Luca Tupputi Schinosa; M Cotrufo
Journal:  Ann Thorac Surg       Date:  1997-04       Impact factor: 4.330

7.  A surgical case for a Carbomedics stuck valve due to thrombotic pannus formation in the mitral position.

Authors:  Naoki Konagai; Hiromi Yano; Susumu Makimura; Kazumasa Nishida; Yasunori Iida; Masahiro Sato; Tatsuhiko Kudo
Journal:  Ann Thorac Cardiovasc Surg       Date:  2005-06       Impact factor: 1.520

8.  Guidelines for the management of patients with valvular heart disease: executive summary. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Management of Patients with Valvular Heart Disease).

Authors:  R O Bonow; B Carabello; A C de Leon; L H Edmunds; B J Fedderly; M D Freed; W H Gaasch; C R McKay; R A Nishimura; P T O'Gara; R A O'Rourke; S H Rahimtoola; J L Ritchie; M D Cheitlin; K A Eagle; T J Gardner; A Garson; R J Gibbons; R O Russell; T J Ryan; S C Smith
Journal:  Circulation       Date:  1998-11-03       Impact factor: 29.690

9.  Differentiating thrombus from pannus formation in obstructed mechanical prosthetic valves: an evaluation of clinical, transthoracic and transesophageal echocardiographic parameters.

Authors:  J Barbetseas; S F Nagueh; C Pitsavos; P K Toutouzas; M A Quiñones; W A Zoghbi
Journal:  J Am Coll Cardiol       Date:  1998-11       Impact factor: 24.094

10.  Long-term results of valve replacement with the St. Jude Medical valve.

Authors:  S Aoyagi; A Oryoji; Y Nishi; K Tanaka; K Kosuga; K Oishi
Journal:  J Thorac Cardiovasc Surg       Date:  1994-12       Impact factor: 5.209

  10 in total
  1 in total

1.  Pannus overgrowth after mitral valve replacement with a Carpentier-Edwards pericardial bioprosthesis.

Authors:  Takeshi Oda; Seiya Kato; Eiki Tayama; Shuji Fukunaga; Hidetoshi Akashi; Shigeaki Aoyagi
Journal:  J Artif Organs       Date:  2009-03-29       Impact factor: 1.731

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.