Influence of anode-filter combinations on image quality and radiation dose in 965 women undergoing mammography.

A C Thilander-Klang1, P H Ackerholm, I C Berlin, N G Bjurstam, S L Mattsson, L G Månsson, C von Schéele, S J Thunberg.   

Abstract

PURPOSE: To evaluate how anode-filter combinations influence image quality in and mean glandular dose to breasts of different thicknesses and compositions.
MATERIALS AND METHODS: Mammograms were obtained with a molybdenum (Mo) anode and a Mo filter at 26 kVp, a Mo anode and a rhodium (Rh) filter at 27 kVp, or a tungsten (W) anode and a Rh filter at 26 kVp in 965 women. One anode-filter-tube voltage combination was used in the right breast and another in the left. The mean glandular dose to each breast was calculated.
RESULTS: Image contrast was highest in the Mo-Mo mammograms. However, depiction of the glandular tissue, pectoral muscle, and skin and subcutis was significantly (P < .001) better with the Mo-Rh and the W-Rh than with the Mo-Mo combination. The average mean absorbed doses to the glandular tissue for W-Rh and Mo-Rh were 50% and 75%, respectively, of that for Mo-Mo.
CONCLUSION: Breast thickness is the most important parameter in selection of an anode-filter-tube voltage combination. Compared with Mo-Mo, both Mo-Rh and W-Rh gave good image quality of the mammary gland and a considerably lower absorbed dose. Mo-Rh-27 kVp is recommended for breast thicknesses of 60 mm or less; W-Rh-26 kVp, for breast thicknesses of greater than 60 mm.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9114087     DOI: 10.1148/radiology.203.2.9114087

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

Review 1.  Task-based measures of image quality and their relation to radiation dose and patient risk.

Authors:  Harrison H Barrett; Kyle J Myers; Christoph Hoeschen; Matthew A Kupinski; Mark P Little
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

2.  Breast tomosynthesis with monochromatic beams: a feasibility study using Monte Carlo simulations.

Authors:  A Malliori; K Bliznakova; I Sechopoulos; Z Kamarianakis; B Fei; N Pallikarakis
Journal:  Phys Med Biol       Date:  2014-08-01       Impact factor: 3.609

Review 3.  Reconstruction of absorbed doses to fibroglandular tissue of the breast of women undergoing mammography (1960 to the present).

Authors:  Isabelle Thierry-Chef; Steven L Simon; Robert M Weinstock; Deukwoo Kwon; Martha S Linet
Journal:  Radiat Res       Date:  2011-10-12       Impact factor: 2.841

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.